FDA Enforcement Class II Terminated

HIgH-Flo Subcutaneous Safety Needle Sets, 2-Needle Set/ 24 Gauge 9mm Needles, REF/Part Numbers RMS22606, RMS22609, RMS22612, RMS12606, RMS22409, RMS32609, RMS32609, RMS12409, RMS12609, RMS62612, RMS62606, RMS12406, RMS22412, RMS12412, RMS42606, RMS62609, RMS52612, RMS42612, RMS32609, RMS22609, RMS22606, RMS42609 & RMS32612.

Recall: Z-0318-2018 · Reported January 17, 2018

Enforcement

Recall Number
Z-0318-2018
Event ID
76569
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Repro-Med Systems, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 17, 2018
Initiation Date
March 10, 2016
Classification Date
January 5, 2018
Termination Date
February 26, 2018
Address
24 Carpenter Rd, N/A, Chester, NY, 10918-1057, United States

Description

HIgH-Flo Subcutaneous Safety Needle Sets, 2-Needle Set/ 24 Gauge 9mm Needles, REF/Part Numbers RMS22606, RMS22609, RMS22612, RMS12606, RMS22409, RMS32609, RMS32609, RMS12409, RMS12609, RMS62612, RMS62606, RMS12406, RMS22412, RMS12412, RMS42606, RMS62609, RMS52612, RMS42612, RMS32609, RMS22609, RMS22606, RMS42609 & RMS32612.

Reason

On 2/24/2016 during post sterilization inspection of part number F180 Precision Flow Rate Tubing, Lot number 1.042/16 the firm found 701 out of 3,195 were rejected due to visual defect, gap in the supplier s manufacturing seal for Part number #317036 5 X 6.5 ST Bag Combo. the firm decided to recall Precision Flow Rate Tubing and RMS HIgH-Flo Subcutaneous Safety Needles Sets due to the sterile barrier packaging being found to have gaps during post-sterilization inspection of one lot of flow tubing, in which 701 out of 3,195 units were rejected for not passing inspection. The firm determined that the issue was large than just the one lot and recalled the product via March 10, 2016 Voluntary Medical Device Corrections and Removal Notification to customers. The recall was conducted without FDA notification.

Code Info

Lot Numbers: 7.069/16, 7.068/16, 7.066/16, 7.065/16, 7.064/16, 7.063/16, 7.062/16, 7.061/16, 7.060/16, 7.059/16, 7.058/16, 7.057/16, 7.056/16, 7.055/16, 7.054/16, 7.053/16, 7.052/16, 7.051/16, 7.050/16, 7.049/16, 7.048/16 & 7.046/16. Expiration: 2019-02

Distribution

Nationwide in US

Quantity

N/A