FDA Enforcement Class II Terminated

Ceres Self-Drilling Variable Screw 4.0 x 12mm The Ceres screws are self-drilling variable screws that are intended to be used with the Ceres-C Stand-Alone Cervical System.

Recall: Z-0316-2018 · Reported January 10, 2018

Enforcement

Recall Number
Z-0316-2018
Event ID
77986
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Amendia, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 10, 2018
Initiation Date
July 7, 2017
Classification Date
January 4, 2018
Termination Date
October 15, 2019
Address
1755 W Oak Pkwy, N/A, Marietta, GA, 30062-2260, United States

Description

Ceres Self-Drilling Variable Screw 4.0 x 12mm The Ceres screws are self-drilling variable screws that are intended to be used with the Ceres-C Stand-Alone Cervical System.

Reason

Incorrect direct marking and color identification for the affected lot of Ceres Self-Drilling Variable Screws. Although the laser mark and color identify the screws as 12 mm in length, some screws in the lot measure 14 mm in length.

Code Info

Model # 12-31-4012, Lot #133039

Distribution

Distributed to states of: AL, FL, GA, NY, TX, and VA.

Quantity