FDA Enforcement Class II Ongoing

intelliPATH Universal HRP Detection Kit, REF: IPK5011G80, containing DAB Buffer, Catalog Number: IPBF5009G20; ONCORE Pro DAB Chromogen, REF: OPRI6056KT180, containing DAB Buffer, Catalog Number: OPRI6055T60; IP DAB Chromogen Kit 80 ml, REF: IPK5010G80, containing DAB Buffer, Catalog Number: IPBF5009G20; DAB kit, REF: OPRI6056KT180, containing DAB Buffer, Catalog Number: OPRI6055T60 Folate Receptor alpha IHC Assay Kit, REF: IPI4006KG10, containing DAB Buffer, Catalog Number: IPBF5009G20

Recall: Z-0315-2025 · Reported November 13, 2024

Enforcement

Recall Number
Z-0315-2025
Event ID
95481
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Biocare Medical, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 13, 2024
Initiation Date
September 25, 2024
Classification Date
November 6, 2024
Address
60 Berry Dr, N/A, Pacheco, CA, 94553-5601, United States

Description

intelliPATH Universal HRP Detection Kit, REF: IPK5011G80, containing DAB Buffer, Catalog Number: IPBF5009G20; ONCORE Pro DAB Chromogen, REF: OPRI6056KT180, containing DAB Buffer, Catalog Number: OPRI6055T60; IP DAB Chromogen Kit 80 ml, REF: IPK5010G80, containing DAB Buffer, Catalog Number: IPBF5009G20; DAB kit, REF: OPRI6056KT180, containing DAB Buffer, Catalog Number: OPRI6055T60 Folate Receptor alpha IHC Assay Kit, REF: IPI4006KG10, containing DAB Buffer, Catalog Number: IPBF5009G20

Reason

Kits containing the DAB chromogen buffer, intended for use in either manual or automated Immunohistochemistry (IHC) staining protocols, may produce weak to no staining, which may lead to test failure requiring the test to be repeated.

Code Info

REF/UDI-DI/Kit Lot(DAB Lot): IPK5011G80/00847627009005/042324(012424A),060523(051823A), 082923(062923A), 011624(072723A), 042324(012424A), 080524(032624A); OPRI6056KT180/00847627024985/020624(111623A),032224(111623A-3); IPK5010G80/00847627008992/071923(051823A), 091923(081023A), 112823(101923A), 100424(011524A), 041624(022624A); OPRI6056KT180/00847627024985/071924(061824A); IPI4006KG10/00847627008985/080323(062923A), 102723(081023A)

Distribution

US Nationwide distribution in the states of MA, AZ, CT, NY, FL, NC, IN, IL, MO, CO, TX, CA, MI, PR, PA, MD, AL, OH, KY.

Quantity

976