FDA Enforcement Class II Terminated

SimpliRED D-Dimer, Product Code DSRK4 The SimpliRED D-dimer assay is a rapid, qualitative test for the detection of cross-linked fibrin degradation products containing the cross-linked D-diner site in human whole blood

Recall: Z-0314-2017 · Reported November 2, 2016

Enforcement

Recall Number
Z-0314-2017
Event ID
74786
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
BBI SOLUTIONS OEM LTD
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
November 2, 2016
Initiation Date
June 1, 2016
Classification Date
October 27, 2016
Termination Date
June 14, 2018
Address
8 Tom McDonald Avenue, Dundee, N/A, United Kingdom

Description

SimpliRED D-Dimer, Product Code DSRK4 The SimpliRED D-dimer assay is a rapid, qualitative test for the detection of cross-linked fibrin degradation products containing the cross-linked D-diner site in human whole blood

Reason

A component of the kit is showing a sensitivity performance detected in stability testing at 15 months of an 18-month shelf life. Qualitative results may be affected. False negative results are possible which may incorrectly influence the clinical diagnostic decision.

Code Info

Lot Numbers containing test reagent lot # 1012-34437): BU183A, BU183B, BU184A, BU185A, BU185B Expiration Date: 31 JUL 2016

Distribution

US distribution to CT and IN

Quantity

420 units (183 units quarantined)