FDA Enforcement
Class II
Ongoing
Brand Name: AESCULAP Product Name: REDUCING CONVERTER 10/12MM TO 5MM Model/Catalog Number: EK087P Software Version: N/A Product Description: REDUCING CONVERTER 10/12MM TO 5MM Component: N/A
Recall: Z-0312-2026
·
Reported November 5, 2025
Enforcement
- Recall Number
- Z-0312-2026
- Event ID
- 97649
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Aesculap Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 5, 2025
- Initiation Date
- September 24, 2025
- Classification Date
- October 24, 2025
- Address
- 3773 Corporate Pkwy, N/A, Center Valley, PA, 18034-8217, United States
Description
Brand Name: AESCULAP Product Name: REDUCING CONVERTER 10/12MM TO 5MM Model/Catalog Number: EK087P Software Version: N/A Product Description: REDUCING CONVERTER 10/12MM TO 5MM Component: N/A
Reason
It was determined there is the potential of silicone fragments detaching from the yellow sealing component.
Code Info
Model No EK087P; UDI-DI: 04046963620059; Lots 52974276, 52952133, 52954633, 52959357, 52964442, 52968887, 52972214, 52977427, 52981208, 52985519, 52988828, 52992514, 52993617, 52997919, 53002094, 53004347, 53006380, 53007625, 53010869, 53013557
Distribution
Worldwide - US Nationwide distribution in the states of AR, AZ, CA, CT, MD, NE, NM, NY, OR, PA, WA, WI and the country of Canada.
Quantity
155 units