BioFire Blood Culture Identification 2 (BCID2) Panel Part No: RFIT-ASY-0147 UDI: 00815381020338 Blood Culture Catalog No./Description: 442023/BD BACTEC Plus Aerobic medium in plastic vials 442021/BD BACTEC Lytic Anaerobic medium in plastic vials 442192//BD BACTEC Plus Aerobic medium in glass vials 442022/BD BACTEC Plus Anaerobic medium in plastic vials 442020/BD BACTEC Peds Plus mediums in plastic vials All lots with an expiry date prior to
Enforcement
- Recall Number
- Z-0312-2022
- Event ID
- 88934
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- BioFire Diagnostics, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 8, 2021
- Initiation Date
- October 11, 2021
- Classification Date
- November 30, 2021
- Termination Date
- August 10, 2023
- Address
- 515 S Colorow Dr, Salt Lake City, UT, 84108-1248, United States
Description
BioFire Blood Culture Identification 2 (BCID2) Panel Part No: RFIT-ASY-0147 UDI: 00815381020338 Blood Culture Catalog No./Description: 442023/BD BACTEC Plus Aerobic medium in plastic vials 442021/BD BACTEC Lytic Anaerobic medium in plastic vials 442192//BD BACTEC Plus Aerobic medium in glass vials 442022/BD BACTEC Plus Anaerobic medium in plastic vials 442020/BD BACTEC Peds Plus mediums in plastic vials All lots with an expiry date prior to
Due to an increased risk of false positive Acinetobacter calcoaceticus- baumannii complex results. The cause may be as a due to non-viable contamination in the blood culture vials.
Blood Culture Identification 2 (BCID2) Panel Part Number: RFIT-ASY-0147 UDI Code: 00815381020338 Blood Culture Media Catalog No./Description: 442023/BD BACTEC Plus Aerobic medium in plastic vials 442021/BD BACTEC Lytic Anaerobic medium in plastic vials 442192//BD BACTEC Plus Aerobic medium in glass vials 442022/BD BACTEC Plus Anaerobic medium in plastic vials 442020/BD BACTEC Peds Plus mediums in plastic vials All BCID2 Panel lots BD Blood Culture Media vials with expiration date of April 30, 2022 or earlier.
Worldwide distribution - U.S. Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY. The Countries of Algeria, Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Bulgaria, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Dominican Rep/, Ecuador, Egypt, France, Georgia, Germany, Greece, Guam, Guatemala, Hong Kong, Hungary, India, Israel Italy, Kuwait, Latvia, Macedonia, Malaysia, Mexico, Netherlands, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Shanghai, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, United Kingdom, Uruguay, Utd/Arab Emir/.
Total 14,140 Kits [11,155 kits (U.S.) and 2,985 kits (O.U.S.)]