PhoenixSpec Calibrator & Kit is used to calibrate the PhoenixSpec & Phoenix AP Instruments. The instruments are then used to measure the turbidity of microbial suspensions equivalent to McFarland standards 0.10 to 4.50.
Enforcement
- Recall Number
- Z-0311-2017
- Event ID
- 75375
- Classification
- Class III
- Status
- Terminated
- Product Type
- Devices
- Firm
- Becton Dickinson & Co.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- November 2, 2016
- Initiation Date
- June 1, 2016
- Classification Date
- October 25, 2016
- Termination Date
- November 8, 2016
- Address
- 7 Loveton Cir, BD Diagnostic Systems, Sparks, MD, 21152-9212, United States
Description
PhoenixSpec Calibrator & Kit is used to calibrate the PhoenixSpec & Phoenix AP Instruments. The instruments are then used to measure the turbidity of microbial suspensions equivalent to McFarland standards 0.10 to 4.50.
BD has determined that the BBL PhoenixSpec Calibrator kits referenced in this recall from the following lots (5364776, 5345796 ) are labeled with an incorrect expiration date (year). The expiration date printed on the individual calibration tubes within the kit are correct but difficult to read on a portion of the tubes.
Part/Catalog Number.: 440911, 441951 Lot / Serial Number. :5364776, 5345796
BD has contacted all 31 US customers and 22 Ex-US sites which received product from the associated catalog numbers. US: AL, CA, CO, KY, MD, MI, MS, NE, NJ, OH, PA, SC, SD, TX, VA, WA and WI. Ex-US: Argentina, Europe, Canada, China, Ecuador, India, Japan, New Zealand, Peru, Singapore, Thailand and Vietnam 4 US government sites were contacted