FDA Enforcement
Class II
Ongoing
Total Thyroxine (T4) ELISA; REF EIA-1781
Recall: Z-0310-2022
·
Reported December 8, 2021
Enforcement
- Recall Number
- Z-0310-2022
- Event ID
- 88949
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- DRG International, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- December 8, 2021
- Initiation Date
- October 18, 2021
- Classification Date
- November 29, 2021
- Address
- 841 Mountain Ave, Springfield, NJ, 07081-3437, United States
Description
Total Thyroxine (T4) ELISA; REF EIA-1781
Reason
Firm has identified two lots of product that do not meet product shelf life, resulting in failed Controls. The clinical impact may be a delay in diagnosis and a consumption of patient samples.
Code Info
lot#: RN-60923 expiration date 12/31/2021; RN-60831 expiration date 12/31/2021
Distribution
Domestic: NY; International: TZ, VN, BD, CO, TT, YE, BO, DE, SV, GE, PY, TG, RU/UA, NI, DO, NG, EG & CM
Quantity
314 total (270 US)