FDA Enforcement Class II Ongoing

Starcam/Camstar; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging

Recall: Z-0303-2026 · Reported October 29, 2025

Enforcement

Recall Number
Z-0303-2026
Event ID
97647
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 29, 2025
Initiation Date
September 12, 2025
Classification Date
October 23, 2025
Address
9, Andrey Sakharov, Haifa, N/A, Israel

Description

Starcam/Camstar; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging

Reason

GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

Code Info

all serial numbers in distribution

Distribution

Worldwide

Quantity

102 units