IBA Proton Therapy System - PROTEUS 235. PTS-11 versions, PTS-12 versions with Orion PPS version OSS6.0. Configured with Orion Patient Positioning System.

Recall: Z-0295-2024 · Reported November 22, 2023

Enforcement

Recall Number
Z-0295-2024
Event ID
93293
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Ion Beam Applications S.A.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 22, 2023
Initiation Date
September 18, 2023
Classification Date
November 14, 2023
Address
Chemin Du Cyclotron 3, N/A, Ottignies-Louvain-La-Neuve, N/A, N/A, Belgium

Description

IBA Proton Therapy System - PROTEUS 235. PTS-11 versions, PTS-12 versions with Orion PPS version OSS6.0. Configured with Orion Patient Positioning System.

Reason

The patient positioning system can perform motions in unexpected directions, creating the potential for crushing of the patient between the patient positioning devices.

Code Info

UDI-DI: (01)05404013801138. Serial numbers (US): SAT133, SAT140, SBF113, SBF128, SBF135 & (OUS): SAT125, SAT127, SAT132, SAT136, SBF107, SBF112, SBF117, SBF124, & SBF125.

Distribution

Worldwide - US Nationwide distribution in the states of Arkansas, Florida, Kansas, and Virginia and the countries of Belgium, France, India, Italy, Netherlands, Spain, Singapore, and Taiwan.

Quantity

5 (US), 9 (OUS)