FDA Enforcement Class II Terminated

Fresenius 132 Gallon Dry Acid Dissolution Unit, P/N 160147. For Use in the Preparation of acid concentrate for hemodialysis.

Recall: Z-0295-2016 · Reported November 25, 2015

Enforcement

Recall Number
Z-0295-2016
Event ID
72484
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Fresenius Medical Care Renal Therapies Group, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 25, 2015
Initiation Date
October 23, 2015
Classification Date
November 18, 2015
Termination Date
March 2, 2021
Address
920 Winter St, N/A, Waltham, MA, 02451-1521, United States

Description

Fresenius 132 Gallon Dry Acid Dissolution Unit, P/N 160147. For Use in the Preparation of acid concentrate for hemodialysis.

Reason

Dissolution units and replacement motors units do not provide full thermal protection for both the Main and Starter coils resulting in the generation of smoke and/or potentially lead to fire.

Code Info

All 132 Gallon Granuflo Dissolution Units and 132 Gallon Dry Acid Dissolution Units with the H218 Motor and Modified H218 Mixer Motors.

Distribution

Nationwide Distribution.

Quantity

348 units