Brilliance CT Big Bore with software version V4.8.0.10421 Model:728244 CT scanner and simulator designed for radiation oncology and therapy.
Enforcement
- Recall Number
- Z-0293-2024
- Event ID
- 93291
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Philips North America Llc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 22, 2023
- Initiation Date
- October 9, 2023
- Classification Date
- November 14, 2023
- Address
- 222 Jacobs St, N/A, Cambridge, MA, 02141-2289, United States
Description
Brilliance CT Big Bore with software version V4.8.0.10421 Model:728244 CT scanner and simulator designed for radiation oncology and therapy.
For oncology users: If the user performed off-set reconstruction on CT device a shift could be observed on the contouring area of the primary and secondary images sets when using image fusion (on TPS) and incorrect coordinates may be sent to TPS system when using absolute patient marking. This may lead to incorrect radiation therapy planning and possible growth or spread of cancer due to incorrect early treatment regimen. For radiology users: If the user performed off-set reconstruction on CT devices, after multiple acquisitions, the Relate Position could be inaccurate. This could lead to incorrect diagnosis and treatment.
UDI-DI: (01)00884838059450 Serial Numbers: 760012 7742 75043 76137 76005 7830 76202 76134 76038 7871 7825
Nationwide Foreign: Austria Belgium Canada China Estonia France Germany Hondorus India Italy Mexico Netherlands Philippines Poland Singapore Spain Switzerland Taiwan
11 units