FDA Enforcement
Class II
Ongoing
Elscint Cardial; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Recall: Z-0290-2026
·
Reported October 29, 2025
Enforcement
- Recall Number
- Z-0290-2026
- Event ID
- 97647
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 29, 2025
- Initiation Date
- September 12, 2025
- Classification Date
- October 23, 2025
- Address
- 9, Andrey Sakharov, Haifa, N/A, Israel
Description
Elscint Cardial; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Reason
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
Code Info
all serial numbers in distribution
Distribution
Worldwide
Quantity
11 units