FDA Enforcement Class II Terminated

RTI Surgical Streamline OCT Occipito-Cervico-Thoracic Spinal Rod, CoCr, 3.2mm (D) x 240mm (L) Product Usage: When intended to promote fusion of the cervical spine and occipitocervico-thoracic junction (occiput-T3), the Streamline OCT Occipito-Cervico-Thoracic System is intended for: degenerative disc disease (as defined by neck or back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, spinal stenosis, fracture/ dislocation, atlanto/axial fracture with instability, occipito-cervical dislocation, deformities or curvature, tumors, pseudoarthrosis, and revision of previous cervical and upper thoracic spine surgery. The occipital bone screws are limited to occipital fixation only. The use of the pedicle screws (standard and high angle) is limited to placement in the upper thoracic spine (T1-T3) in treating thoracic conditions only. The pedicle screws are not intended for use in the cervical spine. The hooks, connectors, and rods are also intended to provide stabilization to promote fusion following reduction of fracture/ dislocation or trauma in the cervical/ upper thoracic (C1-T3) spine.

Recall: Z-0290-2017 · Reported October 26, 2016

Enforcement

Recall Number
Z-0290-2017
Event ID
75242
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.)
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
October 26, 2016
Initiation Date
September 16, 2016
Classification Date
October 19, 2016
Termination Date
June 1, 2017
Address
375 River Park Cir, N/A, Marquette, MI, 49855-1781, United States

Description

RTI Surgical Streamline OCT Occipito-Cervico-Thoracic Spinal Rod, CoCr, 3.2mm (D) x 240mm (L) Product Usage: When intended to promote fusion of the cervical spine and occipitocervico-thoracic junction (occiput-T3), the Streamline OCT Occipito-Cervico-Thoracic System is intended for: degenerative disc disease (as defined by neck or back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, spinal stenosis, fracture/ dislocation, atlanto/axial fracture with instability, occipito-cervical dislocation, deformities or curvature, tumors, pseudoarthrosis, and revision of previous cervical and upper thoracic spine surgery. The occipital bone screws are limited to occipital fixation only. The use of the pedicle screws (standard and high angle) is limited to placement in the upper thoracic spine (T1-T3) in treating thoracic conditions only. The pedicle screws are not intended for use in the cervical spine. The hooks, connectors, and rods are also intended to provide stabilization to promote fusion following reduction of fracture/ dislocation or trauma in the cervical/ upper thoracic (C1-T3) spine.

Reason

RTI Surgical Inc. (RTI) is conducting a voluntary recall on two specific lots of the Streamline¿ OCT Straight Rod, CoCr. 3.2mm (D) x 240mm (L) due to the potential for small cracks to be present on the rods. The small cracks on the rods used in implants could cause reduced strength of these affected implants, and lead to patient injury.

Code Info

Part Number: 23-CC-ROD-240; Lot Number: 234829, UDI Number: (01)0084646802 7179( 10}234829; Lot Number: 246678, UDI Number: {01}00846468027179{10}246678

Distribution

US Nationwide Distribution in the states of: AL, CA, CO, FL, IN, IL, MI, MO, MS, NC, NY, OK, TN, TX, and VA,.

Quantity

66