Manual Diff Dropper, Part Number CHB4001, used for preparing blood samples.

Recall: Z-0287-2020 · Reported November 13, 2019

Enforcement

Recall Number
Z-0287-2020
Event ID
83965
Classification
Class III
Status
Terminated
Product Type
Devices
Firm
Biomedical Polymers, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
N/A
Report Date
November 13, 2019
Initiation Date
September 18, 2019
Classification Date
November 5, 2019
Termination Date
September 15, 2020
Address
16 Chocksett Rd, N/A, Sterling, MA, 01564-2336, United States

Description

Manual Diff Dropper, Part Number CHB4001, used for preparing blood samples.

Reason

The metal cannula of the device either broke and the broken piece left in the blood specimen tube, or slid back into the plastic portion of the device instead of piercing the tube.

Code Info

UPC: (01) 50885380116009 Lot numbers: 20180709A1C, 20180717A1C

Distribution

US nationwide distribution.

Quantity

2,052,000