FDA Enforcement
Class II
Terminated
PT Link Waterbath, Model#PT100. Lab Vision PT Module Plus provides an automated standarized alternative to labor intensive and time consuming dewaxing and epitote retrieval procedures.
Recall: Z-0287-2013
·
Reported November 21, 2012
Enforcement
- Recall Number
- Z-0287-2013
- Event ID
- 53336
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Dako North America Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 21, 2012
- Initiation Date
- May 1, 2008
- Classification Date
- November 13, 2012
- Termination Date
- November 13, 2012
- Address
- 6392 Via Real, Carpinteria, CA, 93013-2921, United States
Description
PT Link Waterbath, Model#PT100. Lab Vision PT Module Plus provides an automated standarized alternative to labor intensive and time consuming dewaxing and epitote retrieval procedures.
Reason
Dako is recalling the PT Link Waterbath because of a possible electrical hazard. Specifically, if the heater pad under the reagent tank becomes wet, it may present an electrical shock hazard when the lid is open.
Code Info
Catalog #: PT-Module
Distribution
Nationwide distribution: USA including states of: AK, CA, CO, CT, DC, FL, GA, HI, IA, IL IN, KS, KY, MA, MD, MI, MN, MO, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, and WA.
Quantity
136 units