FDA Enforcement Class II Ongoing

epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System, SMN 10736515

Recall: Z-0286-2025 · Reported November 13, 2024

Enforcement

Recall Number
Z-0286-2025
Event ID
95504
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Siemens Healthcare Diagnostics Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 13, 2024
Initiation Date
September 24, 2024
Classification Date
November 1, 2024
Address
2 Edgewater Dr, N/A, Norwood, MA, 02062-4637, United States

Description

epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System, SMN 10736515

Reason

Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in one lot of epoc BGEM BUN Test Cards approximately 2 months prior to expiration. The observed maximum negative bias from internal testing is -10 mmol/L. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing. One customer complaint has been associated with this issue. No adverse events have been reported. All affected lots have recently expired before the issuance of the recall.

Code Info

UDI-DI: 00809708121860; Lot Numbers: 08-24092-60 08-24093-60 08-24094-60 08-24095-60 08-24096-60

Distribution

US Nationwide. Canada.

Quantity

1292 units