FDA Enforcement Class I Terminated

EBI Osteogen Implantable Bone Growth Stimulator The OsteoGen Implantable Bone Growth Stimulator is indicated in the treatment of long bone nonunions.

Recall: Z-0286-2019 · Reported November 7, 2018

Enforcement

Recall Number
Z-0286-2019
Event ID
81127
Classification
Class I
Status
Terminated
Product Type
Devices
Firm
Zimmer Biomet, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 7, 2018
Initiation Date
February 19, 2018
Classification Date
November 1, 2018
Termination Date
August 25, 2020
Address
1800 W Center St, N/A, Warsaw, IN, 46580-2304, United States

Description

EBI Osteogen Implantable Bone Growth Stimulator The OsteoGen Implantable Bone Growth Stimulator is indicated in the treatment of long bone nonunions.

Reason

Lack of adequate validation and controls to ensure that product cleanliness parameters were consistently met.

Code Info

All serial numbers expiring prior to March 31, 2019 are affected. Item # 10-1320 - OsteoGen 20/S; Serial Numbers: 520798, 520841, 521079, 521260, 521446, 521886, 520830, 520795, 520789, 521881, 520705, 520799, 520842, 521080, 521261, 521447, 521887, 520831, 520796, 520793, 521875, 520706, 520800, 520843, 521081, 521262, 521448, 521888, 520832, 520797, 520837, 521439, 520708, 520801, 520844, 521082, 521263, 521449, 521889, 520833, 520839, 520838, 521440, 520709, 520802, 520845, 521083, 521264, 521862, 521890, 520834, 520840, 520884, 521253, 520710, 520803, 520846, 521084, 521265, 521863, 520717, 520875, 520886, 520885, 521254, 520711, 520804, 520867, 521086, 521267, 521864, 520718, 520876, 521078, 521255, 520881, 520712, 520805, 520868, 521087, 521268, 521865, 520719, 520878, 521257, 521256, 520882, 520713, 520806, 520869, 521089, 521269, 521866, 520720, 520879, 521259, 521442, 520835, 520714, 520827, 520871, 521090, 521321, 521867, 520721, 520880, 521444, 521443, 520836, 520715, 520828, 520872, 521091, 521322, 521868, 521883, 521873, 521445, 521882, 520787, 520716, 520829, 520873, 521092, 521323, 521869, 521872, 521874, 521884, 521885, 520788, 521094, 521250, 521251, 521093, 521324, 521870, 521871, 521325, 521326, 521437, 521438 and 521252. Item # 10-1328S - OsteoGen - 20/SL; Serial Numbers: 521246, 521367, 521374, 521381, 521388, 521495, 521537, 521387, 521493, 520912, 521247, 521368, 521375, 521382, 521389, 521496, 521538, 521373, 521535, 520913, 521248, 521369, 521376, 521383, 521390, 521497, 521539, 521380, 521492, 520914, 521249, 521370, 521377, 521384, 521490, 521498, 521245, 521366, 521386, 520915, 521364, 521371, 521378, 521385, 521491, 521499, 521536, 521244, 521379, 520916, 521365, and 521372 Item # 10-1320M - OsteoGen - 20/M; Serial Numbers: 520857, 521020, 521284, 521597, 521723, 521934, 522046, 521990, 521932, 520722, 520858, 521021, 521285, 521598, 521724, 521935, 522047, 521991, 521933, 520723, 520859, 521022, 521286, 521599, 521725, 521936, 522048, 521996, 521589, 520724, 520860, 521023, 521287, 521600, 521727, 521937, 522049, 522003, 521590, 520725, 520861, 521024, 521288, 521601, 521728, 521939, 522050, 522004, 521715, 520726, 520862, 521025, 521289, 521602, 521729, 521940, 522051, 522005, 521716, 520727, 520863, 521027, 521296, 521603, 521730, 521941, 522052, 522006, 521926, 520728, 520864, 521030, 521297, 521604, 521731, 521942, 522053, 522007, 521927, 520729, 520865, 521031, 521298, 521605, 521732, 521943, 522054, 522008, 521283, 520730, 520866, 521032, 521299, 521606, 521733, 521944, 522055, 522009, 520847, 520731, 520887, 521033, 521300, 521607, 521734, 521945, 521587, 522010, 521010, 520732, 520888, 521034, 521301, 521608, 521735, 521946, 521165, 522011, 521170, 520733, 520889, 521035, 521302, 521609, 521736, 521947, 521166, 522012, 521588, 520734, 520890, 521036, 521303, 521610, 521737, 521948, 521167, 522013, 521714, 520735, 520891, 521037, 521304, 521611, 521738, 521949, 521168, 522014, 521925, 520736, 520892, 521038, 521305, 521612, 521739, 521950, 521169, 522015, 521282, 520737, 520893, 521039, 521306, 521613, 521740, 521951, 520974, 522036, 521277, 520738, 520894, 521115, 521307, 521614, 521741, 521952, 520975, 522037, 521276, 520739, 520895, 521116, 521308, 521615, 521763, 521953, 520976, 522038, 522121, 520740, 520896, 521117, 521309, 521616, 521764, 521954, 520977, 522039, 522124, 520741, 520897, 521118, 521310, 521617, 521765, 521955, 520979, 522040, 521591, 520742, 520898, 521120, 521311, 521618, 521766, 521956, 521585, 522041, 521592, 520743, 520899, 521121, 521312, 521619, 521767, 521957, 521899, 522042, 521593, 520744, 520900, 521123, 521313, 521620, 521785, 521958, 521924, 522043, 521594, 520745, 520901, 521125, 521314, 521621, 521786, 521959, 521709, 522044, 521595, 520746, 520902, 521126, 521315, 521622, 521787, 521960, 521710, 522045, 521596, 520747, 520903, 521127, 521397, 521624, 521788, 521961, 521711, 521928, 521717, 520748, 520904, 522128, 521398, 521625, 521789, 521962, 521712, 521929, 521718, 520749, 520905, 521129, 521399, 521626, 521790, 521963, 521713, 521930, 521719, 520750, 520906, 521130, 521400, 521627, 521791, 521964, 521583, 521931, 521720, 520751, 520927, 521131, 521403, 521628, 521792, 521965, 521584, 520820, 521721, 520752, 520928, 521132, 521405, 521629, 521793, 521968, 521856, 520821, 521722, 520753, 520929, 521133, 521406, 521630, 521794, 521969, 521857, 520822, 520848, 520754, 520930, 521134, 521408, 521631, 521795, 521970, ,521858, 520823, 520849, 520755, 520931, 521135, 521409, 521632, 521796, 521971, 521859, 520824, 521011, 520756, 520932, 521136, 521410, 521633, 521797, 521972, 521860, 520825, 521012, 520757, 520933, 521137, 521411, 521634, 521798, 521973, 521861, 520826, 521988, 520758, 520934, 521138, 521413, 521635, 521799, 521974, 521891, 521581, 521989, 520759, 520935, 521139, 521416, 521636, 521800, 521975, 521892, 521582, 520850, 520760, 520936, 521140, 521540, 521637, 521842, 521976, 521893, 521707, 520851, 520761, 520937, 521141, 521541, 521638, 521843, 521977, 521698, 521708, 520852, 520762, 520938, 521142, 521544, 521639, 521844, 521978, 521699, 521586, 520853, 520763, 520939, 521143, 521545, 521687, 521845, 521979, 521700, 521280, 520854, 520764, 520940, 521144, 521547, 521688, 521846, 521980, 521701, 521281, 520855, 520765, 520941, 521145, 521563, 521689, 521847, 521981, 521702, 520972, 520856, 520766, 520942, 521146, 521564, 521690, 521848, 521982, 521703, 520973, 521013, 520767, 520943, 521147, 521565, 521691, 521849, 521983, 521704, 521163, 521014, 520768, 520944, 521148, 521566, 521692, 521850, 521984, 521705, 521164, 521015, 520769, 520945, 521149, 521567, 521693, 521851, 521985, 521706, 521894, 521016, 520807, 520946, 521150, 521568, 521694, 521852, 521986, 521572, 521895, 521017, 520808, 520947, 521151, 521569, 521695, 521853, 521896, 521573, 521576, 521018, 520809, 520950, 521152, 521570, 521696, 521854, 521897, 521574, 521577, 521278, 520810, 520951, 521153, 521571, 521697, 521855, 521898, 521575, 521578, 521279, 520811, 520958, 521154, 521579, 520813, 520960, 521156, 520815, 520965, 521158, 520812, 520959, 521155, 521580, 520814, 520961, 521157, 520816, 520967, 521159, 520817, 520968, 521160, 520818, 520970, 521161, 520819, 520971, and 521162 Item # 10-1328M - OsteoGen - 20/ML; Serial Numbers: , 520910, 521340, 521347, 521355, 521361, 521835, 521841, 521354, 521346, 520701, 520911, 521341, 521348, 521356, 521362, 521836, 520909, 521360, 521839, 520703, 521335, 521342, 521351, 521357, 521801, 521837, 520908, 521804, 521803, 520770, 521336, 521343, 521352, 521358, 521802, 521838, 521338, 521840, 521359, 520907, 521337, 521344 and 521353. Item # 10-1318 - OsteoGen - 20/F; Serial Numbers: 521463, 521467, 521471, 521475, 521479, 521514, 521878, 521877, 521474, 521460, 521464, 521468, 521472, 521476, 521510, 521462, 521879, 521513, 521470, 521461, 521465, 521469, 521473, 521477, 521512, 521876, 521880, 521478, and 521466 Item # 10-1348S - OsteoGen - 40/S; Serial Numbers: 520675, 521049, 521233, 521481, 521502, 521516, 521525, 521678, 520673, 520663, 520676, 521050, 521234, 521482, 521503, 521517, 521526, 521679, 520674, 520664, 520677, 521051, 521235, 521483, 521504, 521518, 521527, 521680, 521500, 520665, 521040, 521052, 521236, 521484, 521505, 521519, 521528, 521681, 521501, 520666, 521041, 521053, 521237, 521485, 521506, 521520, 521529, 521682, 521242, 520667, 521042, 521054, 521238, 521486, 521507, 521521, 521530, 521683, 521243, 520668, 521043, 521055, 521239, 521487, 521508, 521522, 521531, 521684, 521058, 520669, 521044, 521056, 521240, 521488, 521509, 521523, 521532, 521685, 521059, 520670, 521045, 521057, 521241, 521489, 521515, 521524, 521533, 521686, 521480, 520671,521046, 520672, 521047, 521048, and 521232. Item # 10-1322S - ,OsteoGen - 40/SL; Serial Numbers: 521114, 521223, 521230, 521776, 521784, 521783, 521818, 521113, 521229, 521000, 521217, 521224, 521231, 521777, 521334, 521782, 521819, 521824, 521222, 521001, 521218, 521225, 521331, 521778, 521221, 521228, 521820, 521817, 521112, 521002, 521219, 521226, 521332, 521780, 521004, 521822, 521821, 521815, 521775, 521003, 521220, 521227, 521333, 521781, 521816, and 521823. Item # 10-1348M - OsteoGen - 40/M; Serial Numbers: 520918, 521062, 521177, 521209, 521645, 521676, 521833, 520784, 520917, 520678, 520919, 521063, 521178, 521210, 521646, 521677, 521834, 520785, 520999, 520679, 520920, 521064, 521179, 521212, 521647, 521742, 521900, 520786, 521060, 520680, 520921, 521065, 521180, 521213, 521648, 521743, 521901, 520781, 521061, 520681, 520922, 521066, 521181, 521215, 521649, 521744, 521902, 520782, 521174, 520682, 520923, 521067, 521182, 521216, 521650, 521745, 521903, 520783, 521175, 520683, 520924, 521068, 521183, 521417, 521651, 521746, 521904, 520996, 521176, 520684, 520925, 521069, 521184, 521418, 521652, 521747, 521905, 520997, 521205, 520685, 520926, 521095, 521185, 521419, 521653, 521748, 521906, 520998, 521206, 520686, 520980, 521096, 521186, 521420, 521654, 521749, 521907, 521171, 521208, 520687, 520981, 521097, 521187, 521421, 521655, 521750, 521908, 521172, 521642, 520688, 520982, 521098, 521188, 521422, 521656, 521751, 521909, 521202, 521643, 520689, 520983, 521099, 521189, 521423, 521657, 521752, 521910, 521203, 521644, 520690, 520984, 521100, 521190, 521424, 521658, 521753, 521911, 521436, 521673, 520691, 520985, 521101, 521191, 521425, 521659, 521754, 521912, 521640, 521674, 520692, 520986, 521102, 521192, 521426, 521660, 521755, 521913, 521670, 521675, 520772, 520987, 521103, 521193, 521427, 521661, 521756, 521914, 521671, 521830, 520773, 520988, 521104, 521194, 521428, 521662, 521757, 521826, 521827, 521831, 520774, 520989, 521105, 521195, 521429, 521663, 521758, 521669, 521828, 521832, 520775, 520990, 521106, 521196, 521430, 521664, 521759, 521201, 521435, 521829, 520776, 520991, 521107, 521197, 521431, 521665, 521760, 520995, 521111, 521672, 520777, 520992, 521108, 521198, 521432, 521666, 521761, 520780, 521173, 521641, 520778, 520993, 521109, 521199, 521433, 521667, 521762, 521668, 521825, 521204, 520779, 520994, 521110, 521200, and 521434. and Item # 10-1322M - OsteoGen - 40/ML. Serial Numbers: 521009, 521327, 521452, 521552, 521771, 521923, 521006, 521007, 521920, 520693, 521070, 521328, 521453, 521553, 521772, 522031, 521396, 521008, 521921, 520694, 521071, 521329, 521454, 521554, 521773, 522032, 521293, 521295, 521561, 520695, 521072, 521330, 521455, 521555, 521774, 522033, 521005, 521451, 521562, 520696, 521073, 521391, 521456, 521556, 521915, 522034, 521919, 521551, 521549, 520697, 521074, 521392, 521457, 521557, 521916, 522035, 521560, 521770, 521550, 520698, 521290, 521393, 521458, 521558, 521917, 521294, 521548, 521922, 521450, 520699, 521291, 521394, 521459, 521559, 521918, 520700, 521292, and 521395.

Distribution

US Nationwide Distribution: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV ; and country of: Australia.

Quantity

4,879 units (2,982 units have been implanted and 537 have been scrapped)