Special Orthopaedic Solution Peri-Acetabular Prosthesis V40 Femoral Extension Component Product Usage: The PAR extension piece is a component of the PAR assembly. The Peri-Acetabular Reconstruction (PAR) endoprosthesis was developed to meet the unique needs of patients who require reconstruction of large acetabular defects after a Type I and II pelvic resection.
Enforcement
- Recall Number
- Z-0286-2013
- Event ID
- 63275
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Stryker Howmedica Osteonics Corp.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 21, 2012
- Initiation Date
- July 16, 2012
- Classification Date
- November 13, 2012
- Termination Date
- September 9, 2015
- Address
- 325 Corporate Drive, Mahwah, NJ, 07430, United States
Description
Special Orthopaedic Solution Peri-Acetabular Prosthesis V40 Femoral Extension Component Product Usage: The PAR extension piece is a component of the PAR assembly. The Peri-Acetabular Reconstruction (PAR) endoprosthesis was developed to meet the unique needs of patients who require reconstruction of large acetabular defects after a Type I and II pelvic resection.
Stryker has become aware that insufficient taper torsional strength may result when a PAR extension piece is used with a forged CoCr stem.
Multiple codes C-2051-7-004 RX00138D FEMORAL PELVIC EXTENSION C-2052-5-004 RX00146D FEMORAL PELVIC EXTENSION C-HM42-7-802 RX4278C 40MM EXT W/FEMALE V40 TAPER C-HM42-7-803 RX4278D 50MM EXT W/FEMALE V40 TAPER C-HM42-7-804 RX4278E 60MM EXT W/FEMALE V40 TAPER C-HM49-4-401 RX4944B PAR 40MM EXT W/FEMALE V40 TAPE C-HM49-4-402 RX4944C PAR 50MM EXT W/FEMALE V40 TAPE C-HM49-4-403 RX4944D PAR 60MM EXT W/FEMALE V40 TAPE C-HM49-9-202 RX4992C PAR EXTENSION C-TAPER 40MM C-HM49-9-203 RX4992D PAR EXTENSION C-TAPER 50MM C-HM49-9-204 RX4992E PAR EXTENSION C-TAPER 60MM C-HM52-1-802 RX5218C PAR FEMORAL EXT V40 TAPER 40MM C-KM10-9-400 HG102AA FEMORAL PELVIC EXTENSION C-KM26-5-204 RX2652E PAR FEM EXT V-40, 40MM C-KM26-5-205 RX2652S PAR FEM EXT V-40 50MM C-KM26-5-206 RX2652T PAR FEM EXT V-40 60MM C-KM28-1-501 RX2815B PAR FEM EXT V-40, 40MM C-KM28-1-502 RX2815C PAR FEM EXT V-40 50MM C-KM28-1-503 RX2815D PAR FEM EXT V-40 60MM C-KM28-4-201 RX2842B PAR FEM EXT V-40, 40MM C-KM28-4-202 RX2842C PAR FEM EXT V-40, 50MM C-KM28-4-203 RX2842D PAR FEM EXT V-40, 60MM C-KM28-6-901 RX2869B PAR FEM EXT V-40, 40MM C-KM28-6-902 RX2869C PAR FEM EXT V-40, 50MM C-KM28-6-903 RX2869D PAR FEM EXT V-40, 60MM C-KM28-7-801 RX2878B PAR FEM EXT V-40, 40MM C-KM28-7-802 RX2878C PAR FEM EXT V-40 50MM C-KM28-7-803 RX2878D PAR FEM EXT V-40 60MM C-KM29-6-101 KM2961B PAR FEM EXT V-40, 50MM C-KM29-6-102 RX2961C PAR FEM EXT V-40, 60MM C-KM29-9-101 RX2991B PAR FEM EXT V-40, 50MM C-KM29-9-102 RX2991C PAR FEM EXT V-40 60MM C-KM30-0-901 RX3009B PAR FEM EXT V-40, 50MM C-KM30-0-902 RX3009C PAR FEM EXT V-40, 60MM C-KM30-1-101 RX3011B PAR FEM EXT V-40, 50MM C-KM30-1-102 RX3011C PAR FEM EXT V-40 60MM C-KM32-0-801 RX3208B PAR FEM EXT V-40, 50MM C-KM32-0-802 RX3208C PAR FEM EXT V-40 60MM C-KM32-9-201 RX3292B 40MM EXT W/FEMALE V40 TAPER C-KM32-9-202 RX3292C 50MM EXT W/FEMALE V40 TAPER C-KM32-9-203 RX3292D 60MM EXT W/FEMALE V40 TAPER C-KM33-6-601 RX3366B 40MM EXT W/FEMALE V40 TAPER C-KM33-6-602 RX3366C 50MM EXT W/FEMALE V40 TAPER C-KM33-6-603 RX3366D 60MM EXT W/FEMALE V40 TAPER C-KM34-8-902 RX3489C 40MM EXT W/FEMALE V40 TAPER C-KM35-4-501 RX3545B 40MM EXT W/FEMALE V40 TAPER C-KM35-4-502 RX3545C 50MM EXT W/FEMALE V40 TAPER C-KM35-4-503 RX3545D 60MM EXT W/FEMALE V40 TAPER C-KM35-5-501 RX3555B 40MM EXT W/FEMALE V40 TAPER C-KM35-5-502 RX3555C 50MM EXT W/FEMALE V40 TAPER C-KM35-5-503 RX3555D 60MM EXT W/FEMALE V40 TAPER C-KM35-9-001 RX3590B 40MM EXT W/FEMALE V40 TAPER C-KM37-5-101 RX3751B 40MM EXT W/FEMALE V40 TAPER C-KM37-5-102 RX3751C 50MM EXT W/FEMALE V40 TAPER C-KM37-5-103 RX3751D 60MM EXT W/FEMALE V40 TAPER C-KM37-7-100 RX3771A 40MM EXT W/FEMALE V40 TAPER C-KM37-7-101 RX3771B 50MM EXT W/FEMALE V40 TAPER C-KM37-7-102 RX3771C 60MM EXT W/FEMALE V40 TAPER C-KM37-7-801 RX3778B 40MM EXT W/FEMALE V40 TAPER C-KM37-7-803 RX3778D 60MM EXT W/FEMALE V40 TAPER C-KM37-8-401 RX3784B 40MM EXT W/FEMALE V40 TAPER C-KM37-8-402 RX3784C 60MM EXT W/FEMALE V40 TAPER C-KM38-1-301 RX3813B 40MM EXT W/FEMALE V40 TAPER C-KM38-1-302 RX3813C 50MM EXT W/FEMALE V40 TAPER C-KM38-1-303 RX3813D 60MM EXT W/FEMALE V40 TAPER C-KM39-1-101 RX3911B 40MM EXT W/FEMALE V40 TAPER C-KM39-1-102 RX3911C 50MM EXT W/FEMALE V40 TAPER C-KM39-1-103 RX3911D 60MM EXT W/FEMALE V40 TAPER C-KM40-1-301 RX4013B 40MM EXT W/FEMALE V40 TAPER C-KM40-1-302 RX4013C 50MM EXT W/FEMALE V40 TAPER C-KM40-1-303 RX4013D 60MM EXT W/FEMALE V40 TAPER C-KM40-6-400 RX4064A 40MM EXT W/FEMALE V40 TAPER C-KM40-7-301 RX4073B 40MM EXT W/FEMALE V40 TAPER C-KM40-7-302 RX4073C 50MM EXT W/FEMALE V40 TAPER C-KM40-9-501 RX4095B 40MM EXT W/FEMALE V40 TAPER C-KM40-9-502 RX4095C 50MM EXT W/FEMALE V40 TAPER C-KM40-9-503 RX4095D 60MM EXT W/FEMALE V40 TAPER C-KM41-3-001 RX4130B 40MM EXT W/FEMALE V40 TAPER C-KM41-3-002 RX4130C 50MM EXT W/FEMALE V40 TAPER C-KM41-3-003 RX4130D 60MM EXT W/FEMALE V40 TAPER C-KM41-6-401 RX4164B 40MM EXT W/FEMALE V40 TAPER C-KM41-6-402 RX4164C 50MM EXT W/FEMALE V40 TAPER C-KM41-6-403 RX4164D 60MM EXT W/FEMALE V40 TAPER C-KM42-7-902 RX4279C 50MM EXT W/FEMALE V40 TAPER C-KM42-7-903 RX4279D 60MM EXT W/FEMALE V40 TAPER C-KM43-0-401 RX4304B 40MM EXT W/FEMALE V40 TAPER C-KM43-0-402 RX4304C 50MM EXT W/FEMALE V40 TAPER C-KM43-0-403 RX4304D 60MM EXT W/FEMALE V40 TAPER C-KM43-1-201 RX4312B 40MM EXT W/FEMALE V40 TAPER C-KM43-1-202 RX4312C 50MM EXT W/FEMALE V40 TAPER C-KM43-1-203 RX4312D 60MM EXT W/FEMALE V40 TAPER C-KM43-1-301 RX4313B 40MM EXT W/FEMALE V40 TAPER C-KM43-1-302 RX4313C 50MM EXT W/FEMALE V40 TAPER C-KM43-1-303 RX4313D 60MM EXT W/FEMALE V40 TAPER C-KM43-2-101 RX4321B 40MM EXT W/FEMALE V40 TAPER C-KM43-4-900 RX4349A 50MM EXT W/FEMALE C-TAPER C-KM43-4-901 RX4349B 60MM EXT W/FEMALE C-TAPER C-KM43-5-901 RX4359B 40MM EXT W/FEMALE V40 TAPER C-KM43-5-902 RX4359C 50MM EXT W/FEMALE V40 TAPER C-KM43-5-903 RX4359D 60MM EXT W/FEMALE V40 TAPER C-KM43-7-901 RX4379B 40MM EXT W/FEMALE V40 TAPER C-KM43-7-902 RX4379C 50MM EXT W/FEMALE V40 TAPER C-KM43-7-903 RX4379D 60MM EXT W/FEMALE V40 TAPER C-KM43-9-501 RX4395B 40MM EXT W/FEMALE V40 TAPER C-KM43-9-502 RX4395C 50MM EXT W/FEMALE V40 TAPER C-KM43-9-503 RX4395D 60MM EXT W/FEMALE V40 TAPER C-KM44-1-701 RX4417B 40MM EXT W/FEMALE V40 TAPER C-KM44-1-702 RX4417C 50MM EXT W/FEMALE V40 TAPER C-KM44-1-703 RX4417D 60MM EXT W/FEMALE V40 TAPER C-KM44-2-701 RX4427B 40MM EXT W/FEMALE V40 TAPER C-KM44-2-702 RX4427C 50MM EXT W/FEMALE V40 TAPER C-KM44-2-703 RX4427D 60MM EXT W/FEMALE V40 TAPER C-KM45-1-801 RX4518B 40MM EXT W/FEMALE V40 TAPER C-KM45-1-802 RX4518C 50MM EXT W/FEMALE V40 TAPER C-KM45-1-803 RX4518D 60MM EXT W/FEMALE V40 TAPER C-KM46-9-301 RX4693B 40MM EXT W/FEMALE V40 TAPER C-KM50-2-202 RX5022C PAR FEM EXT C-TAPER 40MM C-KM50-2-203 RX5022D PAR FEM EXT C-TAPER 50MM C-KM50-2-204 RX5022E PAR FEM EXT C-TAPER 60MM C-KM50-4-102 RX5041C PAR 40MM EXT W/FEMALE V40 TAPE C-KM51-6-202 RX5162C PAR 40MM EXT W/FEMALE V40 TAPE C-M052-9-402 RX5294C PAR FEM EXT V40 40MM C-M053-2-202 RX5322C PAR FEM EXTENSION C-TAPER 40MM C-M053-2-203 RX5322D PAR FEM EXTENSION C-TAPER 50MM C-M053-2-204 RX5322E PAR FEM EXTENSION C-TAPER 60MM C-M054-3-202 RX5432C PAR EXTENTION 50MM V40 TAPER C-M054-3-203 RX5432D PAR EXTENSION 60MM V40 TAPER C-M054-9-902 RX5499C PAR EXTENSION V40 TAPER 40MM C-M055-2-603 RX5526D PAR FEMORAL EXTENSION V40 TAPE C-M055-2-604 RX5526E PAR FEMORAL EXTENSION V40 TAPE C-M055-3-402 RX5534C PAR FEMORAL EXT V40 TAPER 40MM C-M055-3-403 RX5534D PAR FEMORAL EXT V40 TAPER 50MM C-M055-3-404 RX5534E PAR FEMORAL EXT V40 TAPER 60MM C-M055-9-301 RX5593B TI PAR FEMORAL EXT V40 TAPER 4 C-M056-1-102 RX5611C PAR FEMORAL EXT V40 TAPER 40MM C-M056-1-103 RX5611D PAR FEMORAL EXT V40 TAPER 50MM C-M056-1-104 RX5611E PAR FEMORAL EXT V40 TAPER 60MM C-M056-4-202 RX5642C PAR FEM EXT V40 TAPER 40MM C-M056-4-203 RX5642D PAR FEM EXT V40 TAPER 50MM C-M056-4-204 RX5642E PAR FEM EXT V40 TAPER 60MM C-M059-1-202 RX05912C PAR FEM EXT V40 TAPER 40MM C-M062-2-601 RX06226B PAR FEM EXT C-TAPER 40MM C-M062-2-602 RX06226C PAR FEM EXT C-TAPER 50MM C-M062-2-603 RX06226D PAR FEM EXT C-TAPER 60MM C-M062-3-401 RX06234B PERI-ACETABULAR PROSTHESIS FEM C-M062-3-402 RX06234C PERI-ACETABULAR PROSTHESIS FEM C-M062-3-403 RX06234D PERI-ACETABULAR PROSTHESIS FEM C-M062-5-702 RX06257C PAR EXT V40 TAPER 40MM C-M062-5-703 RX06257D PAR EXT V40 TAPER 50MM C-M062-5-704 RX06257E PAR EXT V40 TAPER 60MM C-M062-8-500 RX06285A PAR FEMORAL EXT C-TAPER 60mm C-M062-9-202 RX06292C PAR FEMORAL EXT V40 TAPER 40MM C-M062-9-203 RX06292D PAR FEMORAL EXT V40 TAPER 50MM C-M062-9-204 RX06292E PAR FEMORAL EXT V40 TAPER 60MM C-M063-6-601 RX06366B PAR FEM EXT V40 TAPER 40MM C-M063-6-602 RX06366C PAR FEM EXT V40 TAPER 50MM C-M063-8-702 RX06387C PAR FEM EXT V40 TAPER 40MM C-M064-3-102 RX06431C PAR FEM EXT V40 TAPER 40MM C-M064-3-103 RX06431D PAR FEM EXT V40 TAPER 50MM C-M064-3-104 RX06431E PAR FEM EXT V40 TAPER 60MM C-M066-2-302 RX06623C PAR FEM EXT V40 TAPER 40MM C-M066-2-303 RX06623D PAR FEM EXT V40 TAPER 50MM C-M066-2-304 RX06623E PAR FEM EXT V40 TAPER 60MM C-M066-6-003 RX06660D PAR FEM EXT V40 TAPER 50MM C-M066-6-004 RX06660E PAR FEM EXT V40 TAPER 60MM C-M066-8-602 RX06686C PAR FEM EXT V40 TAPER 50MM C-M066-9-302 RX06693C PAR FEM EXT V40 TAPER 40MM C-M066-9-303 RX06693D PAR FEM EXT V40 TAPER 50MM C-M066-9-304 RX06693E PAR FEM EXT V40 TAPER 60MM C-M067-1-002 RX06710C PAR FEM EXT V40 TAPER 40MM C-M067-1-003 RX06710D PAR FEM EXT V40 TAPER 50MM C-M067-4-402 RX06744C PAR FEMORAL EXTENSION V40 TAPE C-M067-4-403 RX06744D PAR FEMORAL EXTENSION V40 TAPE C-M067-4-404 RX06744E PAR FEMORAL EXTENSION V40 TAPE C-M067-4-702 RX06747C PAR FEMORAL EXTENSION V40 TAPE C-M067-4-703 RX06747D PAR FEMORAL EXTENSION V40 TAPE C-M067-4-704 RX06747E PAR FEMORAL EXTENSION V40 TAPE C-M067-5-105 RX06571C PAR V-40 FEMORAL EXTENSION PI C-M067-5-106 RX06751D PAR V-40 FEMORAL EXTENSION PI C-M067-5-107 RX06751E PAR V-40 FEMORAL EXTENSION PI C-M068-0-902 MHM2LD PAR V-40 FEMORAL EXTENSION PI C-M068-0-903 MHM2LE PAR V-40 FEMORAL EXTENSION PI C-M068-0-904 MHM2LH PAR V-40 FEMORAL EXTENSION PI C-M068-2-402 MHM8YD PAR V40 FEMORAL EXTENSION 40MM C-M068-2-403 MHM8YE PAR V40 FEMORAL EXTENSION 50MM C-M068-6-902 MHN77T PAR V40 TAPER EXTENSION PIECE C-M068-6-903 MHN77V PAR V40 TAPER EXTENSION PIECE C-M068-6-904 MHN77W PAR V40 TAPER EXTENSION PIECE C-M068-7-402 MHN77X PAR V40 TAPER EXTENSION PIECE C-M068-7-403 MHN77Y PAR V40 TAPER EXTENSION PIECE C-M068-7-404 MHN780 PAR V40 TAPER EXTENSION PIECE C-M068-9-902 MHNYT1 PAR V40 TAPER EXTENSION PIECE C-M068-9-903 MHNYT2 PAR V40 TAPER EXTENSION PIECE C-M068-9-904 MHNYT3 PAR V40 TAPER EXTENSION PIECE C-M069-2-802 MHPKKL PAR FEM EXT V40 TAPER 60MM C-M069-4-004 MHRH4X PAR C-TAPER EXTENSION PIECE 60 C-M069-5-102 MHRARE PAR V40 TAPER EXTENSION PIECE C-M069-5-103 MHRARH PAR V40 TAPER EXTENSION PIECE C-M069-5-104 MHRARJ PAR V40 TAPER EXTENSION PIECE C-M069-5-602 MHR3DH PAR V40 TAPER EXTENSION PIECE C-M069-5-603 MHR3E0 PAR V40 TAPER EXTENSION PIECE C-M069-5-604 MHR3DX PAR V40 TAPER EXTENSION PIECE C-M070-0-400 MHTPHK PAR V40 FEMORAL EXTENSION COMP C-M071-4-501 MJMVKP V40 FEMORAL EXTENSION COMPONEN C-M071-4-502 MJM0MW V40 FEMORAL EXTENSION COMPONEN C-M071-4-503 MJM1N5 V40 FEMORAL EXTENSION COMPONEN C-M071-8-001 MJP1N6 V40 FEMORAL EXTENSION COMPONEN C-M071-8-801 MJPVJ4 V40 FEMORAL EXTENSION COMPONEN C-M071-8-802 MJPEPK V40 FEMORAL EXTENSION COMPONEN C-M071-8-803 MJP9T9 V40 FEMORAL EXTENSION COMPONEN C-M072-0-901 MJR51TA C-TAPER FEMORAL EXTENSION COMP C-M072-0-902 MJR20A8 C-TAPER FEMORAL EXTENSION COMP C-M072-0-902 MJRVXNA C-TAPER FEMORAL EXTENSION COMP C-M072-0-903 MJRPR1A C-TAPER FEMORAL EXTENSION COMP C-M078-6-002 MLDP0A PAR EXTENSION PIECE C-TAPER 40 C-M079-0-602 MLE6V3 V40 FEMORAL EXTENSION COMPONEN KM15-3-502 RX1535B FEMORAL EXTENSION COMP 60mm KM15-3-503 RX1535C FEMORAL EXTENSION COMP 80mm KM15-3-504 RX1535D FEMORAL EXTENSION COMP 100mm KM16-9-203 RX1692D 45mm FEMORAL EXTENSION KM16-9-205 RX1692E 60mm FEMORAL EXTENSION KM19-1-302 RX1913C FEMORAL EXT 'C' TAPER - 50mm KM19-1-303 RX1913D FEMORAL EXT 'C' TAPER - 60mm KM19-1-304 RX1913E FEMORAL EXT 'C' TAPER - 80mm KM20-4-102 RX2041C PAR EXTENSION C-TAPER 40mm KM20-4-103 RX2041D PAR EXTENSION C-TAPER 50mm KM20-4-104 RX2041E PAR EXTENSION C-TAPER 60mm KM20-5-701 RX2057B PAR FEM EXTENSION - 60mm KM21-2-402 RX2124C FEM EXTENSION C-TAPER 40mm PAR KM21-2-403 RX2124D FEM EXTENSION C-TAPER 50mm PAR KM21-2-404 RX2124E FEM EXTENSION C-TAPER 60mm PAR KM21-5-302 RX2153C PAR FEMORAL EXTENSION - 40MM KM21-5-303 RX2153D PAR FEMORAL EXTENSION - 50MM KM21-5-304 RX2153E PAR FEMORAL EXTENSION - 60MM KM21-6-802 RX2168C FEM EXTENSION C-TAPER 40mm PAR KM21-6-803 RX2168D FEM EXTENSION C-TAPER 50mm PAR KM21-6-804 RX2168E FEM EXTENSION C-TAPER 60mm PAR KM21-7-702 RX2177C PAR FEM EXTENSION V40 - 40MM KM21-7-703 RX2177D PAR FEM EXTENSION V40 - 50MM KM21-7-704 RX2177E PAR FEM EXTENSION V40 - 60MM KM21-7-901 RX2179E PAR LOCKING SCREW KM21-7-902 RX2179C PAR FEMORAL EXTENSION - 40MM KM21-7-903 RX2179D PAR FEMORAL EXTENSION - 50MM KM21-9-402 RX2194C PAR FEM EXT C-TAPER 40MM KM21-9-403 RX2194D PAR FEM EXT C-TAPER 60MM KM21-9-404 RX2194E PAR FEM EXT C-TAPER 50MM KM21-9-405 RX2194F PAR FEM EXT 40MM KM21-9-406 RX2194G PAR FEM EXT 50MM KM21-9-407 RX2194H PAR FEM EXT 60MM KM22-8-702 RX2287C FEM EXTENSION C-TAPER 40mm PAR KM22-8-703 RX2287D FEM EXTENSION C-TAPER 50mm PAR KM22-8-704 RX2287E FEM EXTENSION C-TAPER 60mm PAR KM23-6-702 RX2367C PAR FEM EXT 40MM V40 KM23-6-703 RX2367D PAR FEM EXT 50MM V40 KM23-6-704 RX2367E PAR FEM EXT 60MM V40 KM23-7-403 RX2374D PAR FEM EXT 50MM V40 KM23-7-404 RX2374E PAR FEM EXT 60MM V40 KM23-7-503 RX2375D PAR FEM EXT 50MM V40 KM23-7-504 RX2375E PAR FEM EXT 60MM V40 KM23-7-902 RX2379C PAR FEM EXT 40MM V40 KM23-7-903 RX2379D PAR FEM EXT 50MM V40 KM23-7-904 RX2379E PAR FEM EXT 60MM V40 KM24-5-902 RX2459C PAR FEM EXTENSION V40 - 40MM KM24-5-903 RX2459D PAR FEM EXTENSION V40 - 50MM KM24-5-904 RX2459E PAR FEM EXTENSION V40 - 60MM KM26-0-902 RX2609C PAR FEM EXT 40MM FEM C-TAPER KM26-0-903 RX2906D PAR FEM EXT 50MM FEM C-TAPER KM26-0-904 RX2906E PAR FEM EXT 60MM FEM C-TAPER KM27-6-402 RX2764C PAR FEM EXT 40MM FEM C-TAPER KM27-6-403 RX2764D PAR FEM EXT 50MM FEM C-TAPER KM27-6-404 RX2764E PAR FEM EXT 60MM FEM C-TAPER KM28-3-700 RX3837A PAR FEM EXT - 80MM - V40 TAPER P-KM11-0-503 KM1105N6 PAR FEMORAL EXTENSION 60MM P-KM11-0-503 KM1105N9 PAR FEMORAL EXTENSION 60MM P-KM11-0-503 KM1105N7 PAR FEMORAL EXTENSION 60MM P-KM11-0-503 KM1105N11 PAR FEMORAL EXTENSION 60MM P-KM11-0-503 KM1105N31 PAR FEMORAL EXTENSION 60MM P-KM11-0-503 KM1105N51 PAR FEMORAL EXTENSION 60MM P-KM11-0-503 KM1105N11 PAR FEMORAL EXTENSION 60MM P-KM11-0-503 KM1105N41 PAR FEMORAL EXTENSION 60MM P-KM11-0-503 KM1105N81 PAR FEMORAL EXTENSION 60MM P-KM11-0-503 KM1105N101 PAR FEMORAL EXTENSION 60MM P-KM11-0-503 X1105Q2 PAR FEMORAL EXTENSION 60MM P-KM11-0-504 KM1105D11 PAR FEMORAL EXTENSION 80MM P-KM11-0-504 KM1105D41 PAR FEMORAL EXTENSION 80MM P-KM11-0-504 KM1105D2 PAR FEMORAL EXTENSION 80MM P-KM11-0-504 KM1105D7 PAR FEMORAL EXTENSION 80MM P-KM11-0-505 KM1105E11 PAR FEMORAL EXTENSION 100MM P-KM11-0-505 KM1105E2 PAR FEMORAL EXTENSION 100MM
Worldwide Distribution - US Nationwide
301 units