FDA Enforcement Class II Ongoing

Cartesion Prime, Digital PET-CT, Model: PCD-1000A/3D.004, PCD-1000A/3D, is a diagnostic imaging system that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems.

Recall: Z-0284-2025 · Reported November 13, 2024

Enforcement

Recall Number
Z-0284-2025
Event ID
95471
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Canon Medical System, USA, INC.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 13, 2024
Initiation Date
September 17, 2024
Classification Date
November 1, 2024
Address
2441 Michelle Dr, N/A, Tustin, CA, 92780-7047, United States

Description

Cartesion Prime, Digital PET-CT, Model: PCD-1000A/3D.004, PCD-1000A/3D, is a diagnostic imaging system that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems.

Reason

When PET-CT system is executing reconstruction, if PET acquisition for another patient is performed (or PET reconstruction for another patient is performed from raw data processing), PET reconstruction control processing may hang up and PET reconstruction may stop as a result, which may necessitate repeat PET/CT examinations.

Code Info

UDI-DI: 04987670103810, Software: V10.8 SP0013, V10.15 SP0006. Serial Numbers: 3DA1982001, 3DA2072002, 3DA2072003, 3DA20X2004, 3DA20Y2005, 3DA2112006, 3DA2132007, 3DA2162008, 3DA2192009, 3DA2242010, 3DA2252011, 3DA2262012, 3DA2272013, 3DA2272014, 3DA2282015, 3DA2292016, 3DA22Y2017, 3DA22Z2018, 3DA2322019, 3DA2322020, 3DA2332021, 3DA2332022, 3DA2342023, 3DA2352024, 3DA2352025, 3DB2362026, 3DB2422029

Distribution

US: NV, AR, MN, IA, IL, NC, LA, KS, MS, CA, AL, MD, CO, IN, TN, AZ, WI, VA

Quantity

27