FDA Enforcement Class II Terminated

IMPAX, PACS, Picture Archive and Communications System, IMPAX CV 7.8.x and IMPAX CV 12.x. Corrects certain demographic information (Patient Name, Patient Medical Record Number and/or Accession Number in previously signed reports that have since changed.

Recall: Z-0283-2016 · Reported November 25, 2015

Enforcement

Recall Number
Z-0283-2016
Event ID
72411
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
AGFA Healthcare Corp.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
November 25, 2015
Initiation Date
September 1, 2015
Classification Date
November 17, 2015
Termination Date
December 15, 2020
Address
10 S Academy St, N/A, Greenville, SC, 29601-2632, United States

Description

IMPAX, PACS, Picture Archive and Communications System, IMPAX CV 7.8.x and IMPAX CV 12.x. Corrects certain demographic information (Patient Name, Patient Medical Record Number and/or Accession Number in previously signed reports that have since changed.

Reason

Customers have experienced IMPAX CV software, specifically, Demographics Manager (DM), not producing correct pdf reports after the reports, at the time of physician signature, were correct.

Code Info

Model # L9M2100

Distribution

Distributed in the states of CA, KY, NH, NY, NC, OH, OR, PA, SC, SD, TN, TX, WA and the country of Canada.

Quantity

19