FDA Enforcement Class II Terminated

BOND Slide Labeler Zebra Printer Model TLP3842 and Model GX4301, for use with BOND-III and BOND-MAX instruments

Recall: Z-0281-2019 · Reported November 7, 2018

Enforcement

Recall Number
Z-0281-2019
Event ID
81036
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Leica Microsystems, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 7, 2018
Initiation Date
September 17, 2018
Classification Date
October 26, 2018
Termination Date
August 21, 2020
Address
1700 Leider Ln, N/A, Buffalo Grove, IL, 60089-6622, United States

Description

BOND Slide Labeler Zebra Printer Model TLP3842 and Model GX4301, for use with BOND-III and BOND-MAX instruments

Reason

The power supply unit for the printer used with the device can potentially overheat or cause a fire hazard because it could contain an improperly-mixed phosphorus compound.

Code Info

Zebra Printer Model TLP3842 - Affected PSU Part Number 808113-001 or P1028888-001 with date codes ranging from 0639XX through 1253XX (YYWWXX, where YY=year and WW=week); Zebra Printer Model GX4301 - Affected PSU Part Number 808101-001 with date codes ranging from 0639XX through 1253XX (YYWWXX, where YY=year and WW=week).

Distribution

Distribution was nationwide. There was also government/military distribution. Foreign distribution was made to Canada.

Quantity

876 devices (in U.S. and Canada)