FDA Enforcement
Class II
Terminated
Ray Station Radiation Therapy Treatment Planning System, Model Numbers 3.0, 3.5 and 4.0
Recall: Z-0281-2014
·
Reported November 20, 2013
Enforcement
- Recall Number
- Z-0281-2014
- Event ID
- 66600
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- RAYSEARCH LABORATORIES AB
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- November 20, 2013
- Initiation Date
- October 23, 2013
- Classification Date
- November 14, 2013
- Termination Date
- February 17, 2017
- Address
- Sveavaegen 9, Stockholm, N/A, Sweden
Description
Ray Station Radiation Therapy Treatment Planning System, Model Numbers 3.0, 3.5 and 4.0
Reason
Ray Search Laboratories found an issue with the dose calculation for Elekta motorized wedges in Ray Station 3.0, 3.5 and 4.0. The error presents when a wedge is used for a beam and the wedge angle differs from 60 degrees.
Code Info
Software version and build numbers 3.0.0.251, 3.5.0.16, 3.5.1.6, 4.0.0.14, 4.0.1.4.
Distribution
US Distribution in the states of: California, Colorado, Delaware, Florida, Maryland, Missouri, New Jersey, New York, Ohio, Pennsylvania, Tennessee, and Virginia.
Quantity
195 units domestically