FDA Enforcement Class II Terminated

HX2 Temp Manage SYS 120V The Terumo HX2 Temperature Management System is a selfcontained, dual channel system for supplying temperature controlled water to heat exchangers used in an extracorporeal circuit.

Recall: Z-0278-2013 · Reported November 21, 2012

Enforcement

Recall Number
Z-0278-2013
Event ID
63344
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Terumo Cardiovascular Systems Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 21, 2012
Initiation Date
October 16, 2012
Classification Date
November 13, 2012
Termination Date
July 11, 2013
Address
6200 Jackson Road, Ann Arbor, MI, 48103-9586, United States

Description

HX2 Temp Manage SYS 120V The Terumo HX2 Temperature Management System is a selfcontained, dual channel system for supplying temperature controlled water to heat exchangers used in an extracorporeal circuit.

Reason

Based on a retrospective review of quality data, AC Control boards used in the manufacture of the HX2 Temperature Management System were identified as being manufactured using the incorrect flux material. Use of the incorrect flux may contribute to premature board failure. A search of TCVS complaints revealed 3 complaints received related to the HX2 failure. Each failure occurred during set up

Code Info

Catalog number: 809810 and serial numbers: 1041, 1083, 1084, and 1101-1116.

Distribution

Nationwide Distribution including CA, HI, KS, LA, MI, MO, OR, PA, UT, and WI

Quantity

19 units