FDA Enforcement Class II Terminated

O-arm 1000 2nd Edition Imaging System, BASE OARM Bl70000027100R SYSTEM lOOV RWK, Model Number Bl70000027100R Product Usage: The O-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The 0-arm Imaging System is compatible with certain Image Guided Surgery Systems.

Recall: Z-0274-2019 · Reported October 31, 2018

Enforcement

Recall Number
Z-0274-2019
Event ID
81153
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Medtronic Navigation, Inc.-Littleton
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 31, 2018
Initiation Date
August 30, 2018
Classification Date
October 25, 2018
Termination Date
June 18, 2020
Address
300 Foster St, N/A, Littleton, MA, 01460-2017, United States

Description

O-arm 1000 2nd Edition Imaging System, BASE OARM Bl70000027100R SYSTEM lOOV RWK, Model Number Bl70000027100R Product Usage: The O-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The 0-arm Imaging System is compatible with certain Image Guided Surgery Systems.

Reason

Software anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version 3.1.7 will be installed on all 2nd edition imaging systems.

Code Info

All Serial Numbers

Distribution

Worldwide - US Distribution in the states: AK, AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI and Puerto Rico. The products were distributed to the following foreign countries: Australia, Austria, Bahrain, Belgium, Brazil, Bulgaria, Canada, Chile, China, Columbia, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Kazakhstan, Korea, Kuwait, Lebanon, Libya, Mexico, Netherlands, New Zealand, Poland, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE, UK, Venezuela.

Quantity

596 total