FDA Enforcement Class II Terminated

Easy Dial Oxygen Regulator, 168708D and 168715D. Eight liter from .0.25 to 8 liters, 15 liter from 0.5 liter to 15 liters of oxygen per minute.

Recall: Z-0272-2016 · Reported November 25, 2015

Enforcement

Recall Number
Z-0272-2016
Event ID
72449
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Precision Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
November 25, 2015
Initiation Date
October 8, 2015
Classification Date
November 13, 2015
Termination Date
October 27, 2016
Address
300 Held Dr, N/A, Northampton, PA, 18067-1150, United States

Description

Easy Dial Oxygen Regulator, 168708D and 168715D. Eight liter from .0.25 to 8 liters, 15 liter from 0.5 liter to 15 liters of oxygen per minute.

Reason

These lots of Oxygen Dial Regulators could have the settings label off by one position. When the indicator arrow points to OFF position, the regulator is actually on the half liter setting (0.50), for the 15 liter. For the 8 liter, instead of being in the OFF position, the regulator will be in the quarter liter setting (0.25).

Code Info

Lot numbers 072215 and 072315.

Distribution

Distributed in the US to the states of PA, MA, MO, MT, ID, MI, MN, ND, KY, NY, NC, CA, AL, AZ, LA, OH, TX, GA, IN, and OK.

Quantity

315