FDA Enforcement Class II Terminated

SCFE (slipped capital femoral epiphysis) Knobs for Driver, 6.5mm and 7.3 mm Orthopedic surgical instrument component, for hip surgery

Recall: Z-0270-2020 · Reported November 13, 2019

Enforcement

Recall Number
Z-0270-2020
Event ID
84017
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Pega Medical Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
November 13, 2019
Initiation Date
October 2, 2019
Classification Date
November 1, 2019
Termination Date
October 26, 2020
Address
1111 Chomedey (A-13) E, N/A, Laval, N/A, N/A, Canada

Description

SCFE (slipped capital femoral epiphysis) Knobs for Driver, 6.5mm and 7.3 mm Orthopedic surgical instrument component, for hip surgery

Reason

There is a potential for the screws to have become loosened during ultrasonic cleaning of the knobs and the screws may come out.

Code Info

Catalogue Numbers: SCF-KNB265, SCF-KNB273, SCF-KNB365, SCF-KNB373; Lot Numbers: 160428-03, 160428-04, A182-05, A182-15, 160603-05, 160603-06, A183-06

Distribution

Nationwide distribution to CA, FL, GA, HI, LA, MA, MO, NC, NY, OK. International distribution to AUSTRALIA, CANADA, DENMARK, GERMANY, ISRAEL, MEXICO, ROMANIA, RUSSIA, UNITED ARAB EMIRATES, UNITED KINGDOM

Quantity

31 devices