FDA Enforcement
Class II
Terminated
L3 Security & Detection System Cabinet X-ray system used for security screening of checked airline baggage.
Recall: Z-0270-2018
·
Reported January 3, 2018
Enforcement
- Recall Number
- Z-0270-2018
- Event ID
- 78658
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- L3 Security & Detection
- Voluntary / Mandated
- FDA Mandated
- Initial Notification
- Letter
- Report Date
- January 3, 2018
- Initiation Date
- October 27, 2017
- Classification Date
- December 22, 2017
- Termination Date
- February 10, 2020
- Address
- 10E Commerce Way, N/A, Woburn, MA, 01801-1044, United States
Description
L3 Security & Detection System Cabinet X-ray system used for security screening of checked airline baggage.
Reason
L3 has determined in airport environments, there have been several cases where a small air gap has developed where the entrance or exit tunnels meet up with the main x-ray scanner tunnel. This a gap in x-ray shielding can result in x-ray emissions that exceed the limit in the federal performance standard for cabinet x-ray systems (see 21 CFR ¿ 1020.40(c)(1)).
Code Info
Model Number : Examiner 3DX 6000 cabinet x-ray system
Distribution
USA (nationwide) Distribution. Owned and operated by the U.S. Transportation Security Administration. Location of individual products is law enforcement sensitive. TSA HQ is located in Crystal City, Virginia.
Quantity
471