FDA Enforcement Class II Ongoing

Alinity m System, REF 08N53-002, For In Vitro Diagnostic Use

Recall: Z-0268-2023 · Reported November 30, 2022

Enforcement

Recall Number
Z-0268-2023
Event ID
91001
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Abbott Molecular, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 30, 2022
Initiation Date
October 12, 2022
Classification Date
November 18, 2022
Address
1300 E Touhy Ave, N/A, Des Plaines, IL, 60018-3315, United States

Description

Alinity m System, REF 08N53-002, For In Vitro Diagnostic Use

Reason

Abbott has identified potential performance issues for the Alinity m System software version 1.6.5: 1) In a unique scenario, the waste chute flapper was found to not open when the Systems Solution drawer was closed and locked. It was found that the software will cause the flapper door to stay in the closed position when the drawer is closed and locked if the flapper had previously been moved out of position. 2) Sample preparation drawer #1 barcode information is used instead of sample prep drawer #2 when the scanned data is not sent to the System Control Center before the next bottle barcode in sample prep drawer#2 is scanned. 3) Under a specific condition, while the system is processing tests and a new test request is made, when the level of the bulk solution is too low to process a test, the software would error stop the system and try to complete all in-process tests. An internal counter is reset which can cause reaction vessels (RVs) already present in an Amplification Detection Unit (or amp detect) to remain there and not be moved to waste so that more RVs can be placed into the amp detect for processing. 4) 4 to 6 replicates of the same auto calibrator orders can be run with 2 different sets of calibrator materials on the same rack. It was discovered during internal testing, when running 4 to 6 replicates of a calibrator, it is possible for the user to use 2 separate lots of material which is not detected by the System Control Center (SCC). If this were to occur, the calibration curve would be created using 2 separate material lots. In normal use scenario, this curve would be made using only 1 lot of material.

Code Info

All serial numbers

Distribution

Worldwide distribution.

Quantity

883 units