FDA Enforcement Class II Ongoing

Western Oxytote DTE Digital VIPR (head only) Model: MNDR-600

Recall: Z-0268-2022 · Reported December 1, 2021

Enforcement

Recall Number
Z-0268-2022
Event ID
88877
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Western/Scott Fetzer Company
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
December 1, 2021
Initiation Date
October 16, 2021
Classification Date
November 19, 2021
Address
875 Bassett Rd, Westlake, OH, 44145-1142, United States

Description

Western Oxytote DTE Digital VIPR (head only) Model: MNDR-600

Reason

VIPR (Valve Integrated Pressure Regulator) - battery rupture, causing the plastic VIPR shroud to break apart with emission of fumes and smoke

Code Info

Units distributed from 05/29/20 to 08/31/21. Digital VIPR units manufactured and distributed after 08/31/21 are outside the scope of this recall and will be recognized by a green and silver control knob sticker (green logo, silver background)

Distribution

CT, MD, MT, OH

Quantity

4974 units