FDA Enforcement
Class II
Terminated
Quadra S Offset Broach Handle The offset broach handle is intended to attach to the broaches of different sizes as the surgeon broaches the femoral bone in preparation for the implantation of the femoral stem.
Recall: Z-0267-2013
·
Reported November 14, 2012
Enforcement
- Recall Number
- Z-0267-2013
- Event ID
- 53337
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Medacta Usa Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 14, 2012
- Initiation Date
- October 13, 2008
- Classification Date
- November 8, 2012
- Termination Date
- November 9, 2012
- Address
- 4725 Calle Quetzal Unit B, N/A, Camarillo, CA, 93012-9101, United States
Description
Quadra S Offset Broach Handle The offset broach handle is intended to attach to the broaches of different sizes as the surgeon broaches the femoral bone in preparation for the implantation of the femoral stem.
Reason
Medacta USA Inc has initiated a recall on the Quadra S Offset Broach Handle because of the potential of the rivet that locks the metallic wire to the lever may become loose during a surgery and make it impossible to lock/unlock the broaches.
Code Info
Ref: 01.10.10.122 (Lot # 085119), Ref: 01.10.10.123 (Lot # 085120), Ref: 01.10.10.124 (Lot # 085244), Ref: 01.10.10.125 (Lot # 085243)
Distribution
Worldwide Distribution -- US including the states of Texas, Nevada and Colorado., and the country of Belgium.
Quantity
12 units