FDA Enforcement Class II Ongoing

Permobil TiLite, Models: Aero Z, and ZRA

Recall: Z-0264-2022 · Reported November 24, 2021

Enforcement

Recall Number
Z-0264-2022
Event ID
88718
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
TiSport, Llc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 24, 2021
Initiation Date
September 24, 2021
Classification Date
November 18, 2021
Address
2701 W Court St, N/A, Pasco, WA, 99301-3911, United States

Description

Permobil TiLite, Models: Aero Z, and ZRA

Reason

Some affected wheelchairs, with transit tie downs, that are used as a passenger seat in a motor vehicle may not be configured in the right orientation to perform as intended if involved in a motor vehicle accident. Some affected wheelchairs can no longer be used as a passenger seat until transit tie downs are replaced with redesigned transit tie downs, verified, and released to the market.

Code Info

Manufactured between January 19, 2017 and June 14, 2021 with the serial number range from M2005054 to 1231112011

Distribution

US Nationwide Distribution: MD, PA, VA, IN, NY, MO, AR, NE, NJ, WA, NM, IL, CA, OH, KY, LA, OK, WI, TX, NC, SD, MN, MA, OR; OUS (Foreign): AUSTRALIA, CANADA, FRANCE, GERMANY, IRELAND, ITALY, UNITED KINGDOM, URUGUAY, NEW ZEALAND, IRELAND, SWEDEN

Quantity

5373