FDA Enforcement Class II Terminated

Samsung GM60A-40S Intended for use in generating radiographic images of human anatomy.

Recall: Z-0263-2016 · Reported November 18, 2015

Enforcement

Recall Number
Z-0263-2016
Event ID
72171
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
NeuroLogica Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
November 18, 2015
Initiation Date
September 14, 2015
Classification Date
November 10, 2015
Termination Date
January 19, 2017
Address
14 Electronics Ave, N/A, Danvers, MA, 01923-1011, United States

Description

Samsung GM60A-40S Intended for use in generating radiographic images of human anatomy.

Reason

When the system is driving over a non-conductive floor and then over a conductive, grounded part, proper control of the drive behavior may not be possible. The system can be stopped by releasing the drive handle or by pressing the emergency switch. Proper behavior is re-enabled when the system is rebooted.

Code Info

Serial Numbers: 50EAM9BG200004X, 50EAM9BG200001A 50EAM9BG200002P, 50EAM9BG200003D, 50EAM9BG200005N, 50EAM9BG200006E, 50EAM9BG200007F

Distribution

Distributed to the states of NC, OH, MA, and MO.

Quantity

7 devices