FDA Enforcement Class II Terminated

0.7/1.3U163CS-36 In combination with below systems MobileArt Evolution, MobileDaRt Evolution Product Usage: The device is a mobile X-ray system in order to take general radiography of patients in the hospital who cannot move and/or of outpatients in emergency.

Recall: Z-0263-2013 · Reported November 14, 2012

Enforcement

Recall Number
Z-0263-2013
Event ID
62271
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Shimadzu Medical Systems
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 14, 2012
Initiation Date
October 28, 2011
Classification Date
November 7, 2012
Termination Date
July 26, 2016
Address
20101 S Vermont Ave, N/A, Torrance, CA, 90502-1328, United States

Description

0.7/1.3U163CS-36 In combination with below systems MobileArt Evolution, MobileDaRt Evolution Product Usage: The device is a mobile X-ray system in order to take general radiography of patients in the hospital who cannot move and/or of outpatients in emergency.

Reason

Shimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standards, 21CFR 1020.30 (m) Beam quality --(1) Half-value layer (HVL), which was revised in 2006. Certain devices might not comply with the requirements for them in some x-ray generating condition after the revision in 2006.

Code Info

S/N of X-ray tube: 64510 66036 66026 66027 71866 70641 72470 72740 72735 73360 73365 2XY0084531 73372 73679 73671 73966 74248 74258 74546 74548 74291 74556 74832 74834 74833 74829 74837 76058 76326 76069 76355 76364 76361 76356 76358 76363 76367 2XY0080666 80666 2XY0080650 2XY0080345 2XY0080649 2XY0080656 2XY0082453 2XY0082456 2XY0082461 2XY0082748 2XY0083039 2XY0075448 2XY0082750 2XY0083070 2XY0082778 2XY0083047 2XY0083058 2XY0083680 2XY0083627 2XY0083633 2XY0083951 2XY0084556

Distribution

US Nationwide Distribution

Quantity

133 units