FDA Enforcement
Class II
Ongoing
The Infinity Acute Care System (IACS) Monitoring Solution with the Standalone Infinity M540 patient monitor. Model nos. MS25510, MS25520, MS26372.
Recall: Z-0258-2020
·
Reported November 6, 2019
Enforcement
- Recall Number
- Z-0258-2020
- Event ID
- 83889
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Draegar Medical Systems, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 6, 2019
- Initiation Date
- September 13, 2019
- Classification Date
- October 31, 2019
- Address
- 6 Tech Dr, Andover, MA, 01810-2434, United States
Description
The Infinity Acute Care System (IACS) Monitoring Solution with the Standalone Infinity M540 patient monitor. Model nos. MS25510, MS25520, MS26372.
Reason
Cybersecurity vulnerabilities may cause device to reboot, lose alarm functionality, and/or lose communication with cockpit and/or the Infinity Network.
Code Info
Software version VG4.1.1/VG4.0.3 and lower
Distribution
US Nationwide distribution.
Quantity
5634