FDA Enforcement Class II Terminated

enGen Track System with TCAutomation Software Version 4.2

Recall: Z-0254-2018 · Reported December 20, 2017

Enforcement

Recall Number
Z-0254-2018
Event ID
77172
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Ortho-Clinical Diagnostics
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 20, 2017
Initiation Date
April 26, 2017
Classification Date
December 14, 2017
Termination Date
August 23, 2021
Address
100 Indigo Creek Dr, N/A, Rochester, NY, 14626-5101, United States

Description

enGen Track System with TCAutomation Software Version 4.2

Reason

A software anomaly with TCA Software V4.2 may potentially cause a delay in reporting of results due to patient samples being routed to an unintended location when samples are reintroduced back onto the enGEN track

Code Info

Serial numbers: 5GV6W2J, 3RHC23J, 7KSC15J, 953252, CXL941J & JG7BFB5J Product Code 952019-EG, 952020-EG, 952055

Distribution

Domestically - GA, IL, MO, NC & NY; Internationally - Canada, England, France, Italy & Spain

Quantity

16