FDA Enforcement Class II Terminated

Sysmex XE-2100 Automated Hematology System; an automated differential cell counter; Sysmex Corporation, Kobe, Japan; Sysmex America, One Nelson C. White Parkway, Mundelein, IL 60060; catalog #983-1341-1 The Sysmex XE-2100 is an automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories. The XE-2100 performs hematology analyses according to the RF/DC detection method, HydroDynamic Focusing (DC Detection), flow cytometry method (using a semiconductor laser), and SLS-hemoglobin method.

Recall: Z-0254-2013 · Reported November 14, 2012

Enforcement

Recall Number
Z-0254-2013
Event ID
62580
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Sysmex America, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 14, 2012
Initiation Date
June 1, 2012
Classification Date
November 6, 2012
Termination Date
August 22, 2016
Address
1 Nelson C White Pkwy, N/A, Mundelein, IL, 60060-9528, United States

Description

Sysmex XE-2100 Automated Hematology System; an automated differential cell counter; Sysmex Corporation, Kobe, Japan; Sysmex America, One Nelson C. White Parkway, Mundelein, IL 60060; catalog #983-1341-1 The Sysmex XE-2100 is an automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories. The XE-2100 performs hematology analyses according to the RF/DC detection method, HydroDynamic Focusing (DC Detection), flow cytometry method (using a semiconductor laser), and SLS-hemoglobin method.

Reason

There is a variation of reticulocyte counts between XE Series (XE-2100, XE-5000) and XT-Series (XT-2000i, XT-4000i) hematology analyzers that may result in a retic counts that are an average of 21% lower on the XE-Series because of a low bias.

Code Info

catalog #983-1341-1, all serial numbers

Distribution

Nationwide distribution: USA including Puerto Rico.

Quantity

1,099 units