FDA Enforcement
Class II
Terminated
STA Satellite Automated Multi-Parameter Analyzer This system is a fully automatic clinical instrument indicated and intended for the performance of tests on human plasma, the results of which aid in the diagnosis of coagulation abnormalities or in monitoring anticoagulant therapy.
Recall: Z-0252-2013
·
Reported November 14, 2012
Enforcement
- Recall Number
- Z-0252-2013
- Event ID
- 63270
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Diagnostica Stago, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- November 14, 2012
- Initiation Date
- June 15, 2012
- Classification Date
- November 6, 2012
- Termination Date
- July 19, 2013
- Address
- 5 Century Dr, N/A, Parsippany, NJ, 07054-4607, United States
Description
STA Satellite Automated Multi-Parameter Analyzer This system is a fully automatic clinical instrument indicated and intended for the performance of tests on human plasma, the results of which aid in the diagnosis of coagulation abnormalities or in monitoring anticoagulant therapy.
Reason
Following a single customer complaint that incorrect INR values were occasionally printed for PT tests, even though the primary units (seconds) were correct, the firm investigated and confirmed the potential for printing incorrect secondary units when using an external printer and the customer printout mode on the STA Satellite¿.
Code Info
K082248 D089976 Ref 58104 units
Distribution
Worldwide Distribution - USA (nationwide) and Canada
Quantity
232 units US; 5 units Canada