FDA Enforcement Class II Terminated

Bio-Stable 5F DL-55CM MST-70 KIT Non-Valved w/ Nit GW, UPN H965458950, Catalog No. 45-895 The Xcela with PASV, BioFlo PICC with ENDEXO Technology, and BioFlo PICC with ENDEXO and PASV Valve Technology are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications, nutrients; the sampling of blood; and for power injection of contrast media. Non-Valve lumens are indicated for central venous pressure monitoring.

Recall: Z-0251-2018 · Reported December 20, 2017

Enforcement

Recall Number
Z-0251-2018
Event ID
76570
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Angiodynamics Inc. (Navilyst Medical Inc.)
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 20, 2017
Initiation Date
November 28, 2016
Classification Date
December 13, 2017
Termination Date
June 12, 2019
Address
10 Glens Falls Tech Park, N/A, Glens Falls, NY, 12801-3864, United States

Description

Bio-Stable 5F DL-55CM MST-70 KIT Non-Valved w/ Nit GW, UPN H965458950, Catalog No. 45-895 The Xcela with PASV, BioFlo PICC with ENDEXO Technology, and BioFlo PICC with ENDEXO and PASV Valve Technology are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications, nutrients; the sampling of blood; and for power injection of contrast media. Non-Valve lumens are indicated for central venous pressure monitoring.

Reason

Navilyst Medical, Inc. (NMI) the manufacturer of the Xcela PICC with PASV, BioFlo PICC with PASV and the BioFlo PICC, conducted this recall to the end user level based on information received from Greatbatch Medical, the manufacturer of the ViaPeel PTFE Peelable Introducer. Greatbatch Medical determined that the products listed in their 11/11/2016 Recall Notification have the potential for the handles to detach from the sheath during use. NMI has confirmed that the affected sheaths, Greatbatch Model Number 10890-006, had been included in packaged Xcela and BioFlo PICC Kits.

Code Info

Lot 4902618

Distribution

Worldwide Distribution: US (Nationwide) and countries of:; Canada, Great Britain, Netherlands, South Australia and Spain.

Quantity

4591 units in total