FDA Enforcement
Class II
Terminated
OneLIF Torque Limiting Adapter, AO to 1/4 SQ, 2.25Nm Limit, Catalog #015-010-0000, a component of NovApproach Spine OneLIF Interbody System instrument kit, Catalog #OneLIF INS.
Recall: Z-0246-2023
·
Reported November 23, 2022
Enforcement
- Recall Number
- Z-0246-2023
- Event ID
- 91016
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Novapproach Spine LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- November 23, 2022
- Initiation Date
- October 12, 2022
- Classification Date
- November 17, 2022
- Termination Date
- April 1, 2026
- Address
- 13900 Tech City Cir Ste 300, N/A, Alachua, FL, 32615-6091, United States
Description
OneLIF Torque Limiting Adapter, AO to 1/4 SQ, 2.25Nm Limit, Catalog #015-010-0000, a component of NovApproach Spine OneLIF Interbody System instrument kit, Catalog #OneLIF INS.
Reason
The torque limiting adapter may not stay locked inside of the ratcheting handle once fully engaged.
Code Info
Torque adapter - Lot #104713, UDI G07001501000000; Kit IDs - 504, 505, 506, 501, 101, and 105.
Distribution
US Nationwide distribution in the states of CA and NY.
Quantity
6 devices