FDA Enforcement Class II Terminated

Rusch Flexi-slip tracheal tube Stylet, Size Ch 6, Sterile, Teleflex Medical.

Recall: Z-0246-2016 · Reported November 18, 2015

Enforcement

Recall Number
Z-0246-2016
Event ID
72037
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Teleflex Medical
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 18, 2015
Initiation Date
August 24, 2015
Classification Date
November 6, 2015
Termination Date
September 25, 2017
Address
2917 Weck Dr, N/A, Research Triangle Park, NC, 27709-0186, United States

Description

Rusch Flexi-slip tracheal tube Stylet, Size Ch 6, Sterile, Teleflex Medical.

Reason

Due to reports of the plastic coating of the stylet splitting and/or breaking off of the stylet.

Code Info

Material #503700-000060, Batch numbers: 13IG21, 13LG06, 14AG04, 14FG06, 14GG01, 14LG06, 15AG28, 15CG24 and 15DG38.

Distribution

Worldwide Distribution -- U.S. Nationwide; International Distribution: Argentina, Austria, Belgium, Belgrade, Cameroon, Canada, Chile, Costa Rica, Croatia, Czech Republic, Ecuador, Denmark, Estonia, Finland, France, French Polynesia, Great Britain, Greece, Guyana, Hong Kong, Ireland, Italy, Jordan, Kuwait, Latvia, Lebanon, Martinique, Netherlands, Panama, Peru, Poland, Portugal, Reunion, Romania, Russia, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, Venezuela, and the United Arab Emirates.

Quantity

233,641 units