FDA Enforcement
Class II
Ongoing
Infection Control (IC) Lifting Strap used with the following Cirrus and Pinnacle Ceiling Lifts: Cirrus 450, Cirrus 600, Cirrus 700, Cirrus 750, and Pinnacle 1000
Recall: Z-0245-2023
·
Reported November 23, 2022
Enforcement
- Recall Number
- Z-0245-2023
- Event ID
- 91000
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Tollos
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 23, 2022
- Initiation Date
- October 24, 2022
- Classification Date
- November 17, 2022
- Address
- 8 Easter Ct Ste J, N/A, Owings Mills, MD, 21117-3294, United States
Description
Infection Control (IC) Lifting Strap used with the following Cirrus and Pinnacle Ceiling Lifts: Cirrus 450, Cirrus 600, Cirrus 700, Cirrus 750, and Pinnacle 1000
Reason
Two customers reported four incidents in which the lifting strap broke while lifting a patient. If the IC strap fails during use, the patient may fall causing injury to them or the caregiver(s).
Code Info
UDI-DI 00817956021 (not yet uploaded to GUDID)
Distribution
Worldwide - US Nationwide distribution and the countries of Canada and United Arab Emirates.
Quantity
1273 US; 70 OUS