Torque Limiting Adaptor, AO Connect, 4 N-m, part #027-203, reusable, a component of the Bodycad Fine Osteotomy Implant and Disposable Instrumentation Tray, part #025-004.
Recall: Z-0244-2022
·
Reported November 24, 2021
Enforcement
- Recall Number
- Z-0244-2022
- Event ID
- 88878
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Laboratoires Bodycad, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 24, 2021
- Initiation Date
- July 2, 2021
- Classification Date
- November 17, 2021
- Termination Date
- January 4, 2023
- Address
- 2035 Du Haut-Bord St, N/A, Quebec, N/A, N/A, Canada
Description
Torque Limiting Adaptor, AO Connect, 4 N-m, part #027-203, reusable, a component of the Bodycad Fine Osteotomy Implant and Disposable Instrumentation Tray, part #025-004.
Reason
There is a risk of breakage of the torque limiting adaptor if used counterclockwise and it must not be used with a power tool due to no test data validating this use.
Code Info
Torque Limiting Adaptor - Lot number SB000430. Fine osteotomy tray - Lot numbers SB000251, SB000252, SB000253, SB000254, SB000259, SB000261, SB000263, SB000264, SB000272, and SB000273.
Distribution
Distribution was made to CO, MI, and NY. Foreign distribution was made to Canada and the United Kingdom. There was no military/government distribution.
Quantity
10 in the U.S.