FDA Enforcement
Class II
Terminated
A&E Medical Thorecon Plating System Kits Model No. 94-1300-04-S. For fracture fixation or reconstruction of the sternum.
Recall: Z-0243-2020
·
Reported November 6, 2019
Enforcement
- Recall Number
- Z-0243-2020
- Event ID
- 83942
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Alto Development Corp
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 6, 2019
- Initiation Date
- September 10, 2019
- Classification Date
- October 29, 2019
- Termination Date
- May 12, 2020
- Address
- 5206 Asbury Rd, Wall Township, NJ, 07727-3609, United States
Description
A&E Medical Thorecon Plating System Kits Model No. 94-1300-04-S. For fracture fixation or reconstruction of the sternum.
Reason
Some Thorecon disposable tensioners supplied in different Plating System Kits exhibit a molding defect which occludes the through-channel for cable tensioning, rendering the tensioner inoperable.
Code Info
327703, 327704, 330464, 344282
Distribution
Distributed to AL, CA, FL, GA, IL, LA, MO, OK, MS, NM, OK, SC, TX, UT, VT, WA, and WI.
Quantity
209 total