XCELA PASV 5F DL 55CM IR-145 KIT PG, UPN H965457330, Catalog No. 45-733 The Xcela with PASV, BioFlo PICC with ENDEXO Technology, and BioFlo PICC with ENDEXO and PASV Valve Technology are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications, nutrients; the sampling of blood; and for power injection of contrast media. Non-Valve lumens are indicated for central venous pressure monitoring.
Enforcement
- Recall Number
- Z-0243-2018
- Event ID
- 76570
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Angiodynamics Inc. (Navilyst Medical Inc.)
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 20, 2017
- Initiation Date
- November 28, 2016
- Classification Date
- December 13, 2017
- Termination Date
- June 12, 2019
- Address
- 10 Glens Falls Tech Park, N/A, Glens Falls, NY, 12801-3864, United States
Description
XCELA PASV 5F DL 55CM IR-145 KIT PG, UPN H965457330, Catalog No. 45-733 The Xcela with PASV, BioFlo PICC with ENDEXO Technology, and BioFlo PICC with ENDEXO and PASV Valve Technology are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications, nutrients; the sampling of blood; and for power injection of contrast media. Non-Valve lumens are indicated for central venous pressure monitoring.
Navilyst Medical, Inc. (NMI) the manufacturer of the Xcela PICC with PASV, BioFlo PICC with PASV and the BioFlo PICC, conducted this recall to the end user level based on information received from Greatbatch Medical, the manufacturer of the ViaPeel PTFE Peelable Introducer. Greatbatch Medical determined that the products listed in their 11/11/2016 Recall Notification have the potential for the handles to detach from the sheath during use. NMI has confirmed that the affected sheaths, Greatbatch Model Number 10890-006, had been included in packaged Xcela and BioFlo PICC Kits.
Lot 4899431
Worldwide Distribution: US (Nationwide) and countries of:; Canada, Great Britain, Netherlands, South Australia and Spain.
4591 units in total