FDA Enforcement Class II Ongoing

VITROS Chemistry Products Lipase Slides, Product Codes 1668409 (contains 60 Slides per Cartridge) and 8297749 (contains 18 slides per cartridge)

Recall: Z-0236-2024 · Reported November 15, 2023

Enforcement

Recall Number
Z-0236-2024
Event ID
93238
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Ortho-Clinical Diagnostics, INc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 15, 2023
Initiation Date
September 12, 2023
Classification Date
November 6, 2023
Address
513 Technology Blvd, Rochester, NY, 14626-3601, United States

Description

VITROS Chemistry Products Lipase Slides, Product Codes 1668409 (contains 60 Slides per Cartridge) and 8297749 (contains 18 slides per cartridge)

Reason

Products 1668409 and 8297749 may contain VITROS Lipase Slides with an abnormality which could cause -46% to +86% bias in lipase results depending on how the slides are used. Falsely low levels could result in a missed or delayed diagnosis of acute pancreatitis and may lead to diagnostic confusion in the setting of acute abdominal symptoms. Falsely elevated levels would not result in patient harm but may cause diagnostic confusion and a delay in accurate diagnosis of a patient with abdominal pain.

Code Info

Product Code 1668409: UDI-DI: 10758750009510, Affected Lots: 5905-3348-1557; Product Code 8297749: UDI-DI: 10758750004683, Affected Lots: 5905-3348-2899

Distribution

US States: CA, CO, FL, GA, IL, IN, KS, NC, NY, TN, TX, VT. Canada, France, India, Japan, South Korea

Quantity

342 units