Accutorr 7 Vital Signs Monitor with NIBP valve assembly board, p/n 115-017679-00, containing software version 8.1.5 - Product Usage: The monitor is intended for monitoring physiologic parameters, including Pulse Oximetry (SpO2), Pulse Rate (PR), Non Invasive Blood Pressure (NIBP). Temperature (TEMP) and Carbon Dioxide (CO2) on adult, pediatric, and neonatal patients in professional healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians
Enforcement
- Recall Number
- Z-0236-2020
- Event ID
- 83979
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Mindray DS USA, Inc. dba Mindray North America
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 6, 2019
- Initiation Date
- October 1, 2019
- Classification Date
- October 29, 2019
- Termination Date
- August 6, 2021
- Address
- 800 Macarthur Blvd, Mahwah, NJ, 07430-2001, United States
Description
Accutorr 7 Vital Signs Monitor with NIBP valve assembly board, p/n 115-017679-00, containing software version 8.1.5 - Product Usage: The monitor is intended for monitoring physiologic parameters, including Pulse Oximetry (SpO2), Pulse Rate (PR), Non Invasive Blood Pressure (NIBP). Temperature (TEMP) and Carbon Dioxide (CO2) on adult, pediatric, and neonatal patients in professional healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians
NIBP valve assembly board subject to this recall includes enhancements that are not included in the current marketing clearance:1. Anti-motion feature 2. NIBP hose blockage identification function
NIBP valve assembly board, p/n 115-017679-00, containing software version 8.1.5
US Nationwide distributions.
107 units