VariSoft infusion set, Model Number 1002828
Enforcement
- Recall Number
- Z-0233-2024
- Event ID
- 93269
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Unomedical A/S
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- November 29, 2023
- Initiation Date
- October 11, 2023
- Classification Date
- November 20, 2023
- Address
- Aaholmvej 1 - 3, OSTED, LEJRE, N/A, N/A, Denmark
Description
VariSoft infusion set, Model Number 1002828
The Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily from the insulin set than expected, requiring less force to disconnect than intended. Disconnection, when not detected, for example during sleep, leads to missed basal dosing. which in turn can result in elevated blood glucose and ketone level (nocturnal hyperglycemia). Diabetic ketoacidosis.hyperglycemia). Diabetic ketoacidosis.
GTIN 05705244018730, Serial Numbers: 5388367, 5388376; GTIN 05705244018723, Serial Numbers: 5388367. 5388376
US Nationwide distribution in the state of CA.
1679 units