FDA Enforcement Class II Ongoing

Quest Spectrum 3 Fluorescent Imaging System Model/Catalog Number: ART-SYS-0029, ART-SYS-0030 The Quest Spectrum3 is a fluorescent imaging system used in capturing and viewing fluorescent images for the visual assessment of blood flow for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used during plastic and reconstructive surgeries.

Recall: Z-0232-2025 · Reported November 6, 2024

Enforcement

Recall Number
Z-0232-2025
Event ID
95423
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Olympus Corporation of the Americas
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 6, 2024
Initiation Date
September 24, 2024
Classification Date
October 25, 2024
Address
3500 Corporate Pkwy, N/A, Center Valley, PA, 18034-8229, United States

Description

Quest Spectrum 3 Fluorescent Imaging System Model/Catalog Number: ART-SYS-0029, ART-SYS-0030 The Quest Spectrum3 is a fluorescent imaging system used in capturing and viewing fluorescent images for the visual assessment of blood flow for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used during plastic and reconstructive surgeries.

Reason

Does not meet the requirements for electromagnetic compatibility (EMC) of the standard IEC 60601-1-2. Electromagnetic interference occurring during a procedure may cause the device to malfunction requiring the replacement of the device. This could potentially result in a prolonged procedure and/or require additional medical intervention

Code Info

UDI-DI: 08719324487047 Lot Code: Model: ART-SYS-0029, Serial: ART-SYS-2022-F04, ART-SYS-2022-A06, ART-SYS-2022-F05; Model: ART-SYS-0030; No distribution in the USA

Distribution

OH

Quantity

3 units