ABL800 analyzer with FLEXQ module. Device intended for in vitro testing of samples of whole blood for the parameters pCO2, cK+, cNa+, cCa2+, cCl , cGlu, cLac, cCrea, ctBil, and co-oximetry parameters (ctHb, sO2, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb and FHbF) - in vitro testing of samples of expired air for the parameters pO2 and pCO2. It is also used for in vitro testing of pleura samples for the pH parameter
Enforcement
- Recall Number
- Z-0231-2018
- Event ID
- 78445
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Radiometer America Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 20, 2017
- Initiation Date
- October 24, 2017
- Classification Date
- December 8, 2017
- Termination Date
- October 10, 2023
- Address
- 250 S Kraemer Blvd, N/A, Brea, CA, 92821-6232, United States
Description
ABL800 analyzer with FLEXQ module. Device intended for in vitro testing of samples of whole blood for the parameters pCO2, cK+, cNa+, cCa2+, cCl , cGlu, cLac, cCrea, ctBil, and co-oximetry parameters (ctHb, sO2, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb and FHbF) - in vitro testing of samples of expired air for the parameters pO2 and pCO2. It is also used for in vitro testing of pleura samples for the pH parameter
Due to misinterpretation of the barcode by the scanner, when the registration receipt barcode is scanned by the analyzer , a result from a different patient will be printed or displayed on the analyzer.
All serial numbers
USA (nationwide) to: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY. Distributed internationally to Canada.
1,749 analyzers